Drager Fabius

Flow Sensor Calibration Failure or No Flow Reading

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius

What This Guide Helps With

Helps isolate sensor installation, contamination, connection, circuit, calibration-condition, or external accessory problems when flow measurement is absent or calibration will not complete.

Step-by-Step Troubleshooting

1. Protect the Patient

Do not troubleshoot an unreliable flow measurement system while a patient depends on the affected monitoring or ventilation function. Use an alternate verified anesthesia machine or monitoring method as clinically appropriate.

Expected outcome: Patient care continues safely while the Fabius is evaluated.

2. Confirm the Exact Failure

Determine whether the flow value is absent, fixed, erratic, obviously incorrect, or whether a calibration attempt fails.

Reproduce the problem off-patient under an approved test configuration.

Expected outcome: The specific flow-sensor symptom is confirmed.

3. Inspect Sensor Installation

Inspect the externally accessible flow sensor and associated breathing-system components. Confirm correct orientation, seating, and assembly.

Look for:

Expected outcome: The flow sensor is installed correctly. If reseating restores a stable reading, proceed to verification.

4. Check for Moisture or Contamination

Inspect the accessible sensor pathway for condensation, fluid, debris, or contamination that could interfere with measurement.

Handle and clean only in accordance with approved manufacturer procedures. Do not improvise cleaning methods.

Expected outcome: The sensor pathway is clean and dry enough for valid operation. If correcting an approved external contamination issue restores measurement, continue to final testing.

5. Inspect External Sensor Connections

Check associated electrical connectors, pneumatic tubing, or accessible sensor interfaces as applicable. Look for looseness, bent contacts, strain, or visible damage.

Expected outcome: All external sensor connections are secure and undamaged.

6. Verify Calibration Conditions

Confirm the breathing system is assembled correctly and placed in the proper non-patient condition required for the approved calibration process.

Do not use undocumented calibration sequences or service menus.

Expected outcome: Calibration is attempted only under valid conditions. If calibration succeeds, proceed to functional verification.

7. Substitute a Known-Good Compatible Sensor

If permitted and available, test with a compatible known-good flow sensor or approved external component.

Expected outcome: If the known-good component calibrates and measures correctly, the original sensor or associated external component is identified as defective and can be replaced.

8. Compare Flow With Test Equipment

Use approved anesthesia or ventilator test equipment to confirm whether gas flow is actually present and whether displayed flow follows expected changes.

Expected outcome: The distinction between a true flow-delivery problem and a measurement problem is established.

9. Perform Final Verification

After correction, complete the applicable calibration and functional checks. Verify stable flow readings, tidal-volume behavior, ventilation, and alarm response with a test lung.

Expected outcome: Flow measurement is stable and appropriate throughout testing. If so, troubleshooting is complete.

10. Escalate if Calibration Still Fails

If the sensor is correctly installed, external connections are intact, known-good components do not resolve the issue, and calibration still fails, stop external troubleshooting.

Expected outcome: The machine is removed from service and escalated for internal diagnostic evaluation.

If the Problem Persists

External installation, contamination, sensor, circuit, and calibration-condition causes have been ruled out. Remaining possibilities may involve internal signal processing, pneumatic sensing, calibration storage, interface electronics, or other service-level functions.

The device should be:

Complete applicable ventilation, flow, volume, alarm, and safety checks before return to service.

Clinical Use Tip

Do not rely on displayed tidal volume when the flow-measurement system is known to be unreliable; move the patient to a verified monitoring and ventilation system first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported no flow reading and an unsuccessful flow-sensor calibration during setup."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external flow sensor was not fully seated in the breathing-system assembly."

Resolution

What action was taken.

Example:
"The sensor was correctly reseated, calibration completed successfully, and flow and ventilation performance were verified with approved test equipment."

Helpful Details to Include (If Known)

Final Thought

A failed flow calibration does not automatically indicate an internal fault. Confirm installation, cleanliness, connections, calibration conditions, and known-good external components first, then escalate with objective test results when the problem remains.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide