Drager Fabius GS

Alarm: APNOEA FLOW !!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

APNOEA FLOW alarm indicating low or absent tidal volume (<20 mL), suggesting apnea, circuit disconnection, or leak impacting ventilation.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Patient Connection

4. Inspect the Breathing Circuit for Leaks or Disconnections

5. Check Ventilator Mode and Settings

6. Assess Gas Flow and Supply

7. Check Sensors and Flow Measurement

8. Perform a System Leak Test

9. Stop When Resolved

If the Problem Persists

If all external causes have been ruled out and the alarm continues, an internal flow sensor or ventilator fault is likely.

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot ventilation alarms on an active patient. Always move the patient to a backup device before continuing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"APNOEA FLOW alarm active with low tidal volume during ventilation."

Cause

What was observed during troubleshooting.

Example:
"Leak found at Y-piece connection causing insufficient tidal volume delivery."

Resolution

What action was taken.

Example:
"Re-secured circuit connection and verified normal ventilation and alarm resolution."

Helpful Details to Include (If Known)

Final Thought

Apnea-related alarms directly indicate loss of ventilation and must be addressed immediately. Always start with simple external checks, prioritize patient safety, and escalate when necessary.

That is successful troubleshooting.