Drager Fabius GS

Alarm: APNOEA PRESSURE !!!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

APNOEA PRESSURE !!! alarm indicating airway pressure is not being generated, resulting in stopped ventilation due to disconnection, leak, or ventilator failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Breathing Circuit for Disconnection

4. Check for Circuit Leaks

5. Inspect CO₂ Absorber and Components

6. Check Ventilator Settings and Mode

7. Verify Gas Supply

8. Perform Manual Ventilation Test

9. Isolate Ventilator vs Circuit Issue

If the Problem Persists

If all external causes have been ruled out and the alarm continues, this indicates internal ventilator failure or control issue.

Clinical Use Tip

Never troubleshoot ventilation alarms while the patient depends on the machine. Always secure ventilation first using manual or alternate equipment.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit alarming APNOEA PRESSURE !!! with no ventilation and no airway pressure observed during use."

Cause

What was observed during troubleshooting.

Example:
"Leak or disconnection in breathing circuit or suspected ventilator failure after verifying connections, leak test, and gas supply."

Resolution

What action was taken.

Example:
"Secured circuit connections, performed leak test, verified gas supply and settings. Alarm persisted → removed from service and sent for repair."

Helpful Details to Include (If Known)

Final Thought

Apnea pressure alarms indicate failure to generate airway pressure—most commonly due to disconnections or leaks. Always start with simple external checks and escalate appropriately to protect patient safety.

That is successful troubleshooting.