Drager Fabius GS

Alarm: APNOEA VENTILATION !!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

APNOEA VENTILATION !! alarm indicating stopped breathing or failed ventilation, typically caused by circuit disconnection, leaks, or incorrect ventilation settings leading to loss of effective ventilation.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Alarm and Behavior

3. Check Breathing Circuit Connections

4. Check for Leaks in the Breathing System

5. Verify Gas Flow and Ventilator Operation

6. Evaluate Ventilation Mode and Settings

7. Check Pressure Support Settings

8. Observe Breath Triggering

9. Confirm Resolution

If the Problem Persists

If all external causes such as disconnections, leaks, gas supply, and settings have been ruled out:

Clinical Use Tip

Never troubleshoot ventilation alarms on an active patient. Always move the patient to a backup device before attempting troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit alarming APNOEA VENTILATION !!; ventilator not delivering effective breaths."

Cause

What was observed during troubleshooting.

Example:
"Loose breathing circuit connection causing loss of ventilation and apnea alarm."

Resolution

What action was taken.

Example:
"Reconnected circuit, verified no leaks, confirmed proper ventilation and cleared alarm."

Helpful Details to Include (If Known)

Final Thought

APNOEA VENTILATION alarms indicate a critical loss of ventilation. Focus on simple external causes first, prioritize patient safety, and escalate appropriately if ventilation cannot be restored.

That is successful troubleshooting.