Drager Fabius GS

Alarm: CHECK APL VALVE !!!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

Alarm CHECK APL VALVE !!! indicating unintended expiratory flow during inspiration, which can cause improper ventilation, system leaks, or APL/valve control malfunction affecting patient safety.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Breathing Circuit for External Causes

4. Inspect the APL Valve Setting and Function

5. Check Expiratory Valve Assembly

6. Inspect Expiratory Control Line / Hose

7. Check Flow Sensor Condition

8. Perform Flow Sensor Calibration

9. Re-test the System

If the Problem Persists

If external causes have been ruled out, the issue may involve internal valve control, pneumatics, or flow measurement systems.

Clinical Use Tip

Never troubleshoot breathing system or valve-related alarms while connected to a patient. Always switch to a backup ventilation method first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit alarming CHECK APL VALVE !!! during ventilation with abnormal flow behavior."

Cause

What was observed during troubleshooting.

Example:
"Contamination found in expiratory valve preventing proper sealing."

Resolution

What action was taken.

Example:
"Cleaned/reseated expiratory valve, verified circuit integrity, performed flow sensor calibration, alarm cleared."

Helpful Details to Include (If Known)

Final Thought

This alarm indicates a critical ventilation integrity issue. Always start with simple external causes, but escalate when system control components are involved to ensure patient safety.

That is successful troubleshooting.