Drager Fabius GS

Alarm: CHECK BATTERY !

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

Alarm CHECK BATTERY ! indicating 0% backup battery charge, resulting in no battery backup during power loss and risk of ventilation interruption.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check AC Power Source

Inspect Power Cord and Connections

Allow Battery Charging Time

Check for Immediate Power Loss on Unplug

Inspect External Indicators and Physical Condition

Check External Fuse (If Accessible)

If the Problem Persists

If the battery remains at 0% and the alarm continues after verifying external power and connections, external causes have been ruled out. This indicates a failed battery or internal charging circuit issue.

Clinical Use Tip

Never troubleshoot or rely on a device with a failed battery while connected to a patient. Always move the patient to a backup device before continuing evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit displaying CHECK BATTERY ! alarm. Battery indicated 0% and no backup power available."

Cause

What was observed during troubleshooting.

Example:
"Battery failed to charge despite verified AC power; likely depleted battery or internal charging circuit failure."

Resolution

What action was taken.

Example:
"Removed from service, labeled out of service, and sent for battery replacement. Verified immediate shutdown when unplugged."

Helpful Details to Include (If Known)

Final Thought

Battery-related alarms directly impact patient safety and device reliability. Always verify power sources first, confirm charging behavior, and escalate promptly when backup capacity is lost.

That is successful troubleshooting.