Drager Fabius GS

Alarm: CONTINUOUS PRESSURE !!!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

CONTINUOUS PRESSURE !!! alarm indicating sustained elevated airway pressure for more than 15 seconds, which can lead to unsafe ventilation conditions and potential patient harm.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Breathing Circuit for Obstruction or Kinking

4. Check Patient Interface and Airway

5. Check APL Valve (ManSpont Mode)

6. Check Fresh Gas Flow Settings

7. Check Pressure Alarm Limits

8. Check for Valve or Internal Flow Restriction Indicators

If the Problem Persists

If all external causes have been ruled out and airway pressure remains elevated:

Clinical Use Tip

Never troubleshoot high airway pressure alarms on an active patient. Always move the patient to a backup device before investigating.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CONTINUOUS PRESSURE !!! alarm with sustained high airway pressure during ventilation."

Cause

What was observed during troubleshooting.

Example:
"Breathing circuit kinked at Y-piece causing obstruction and elevated pressure."

Resolution

What action was taken.

Example:
"Repositioned breathing circuit, verified normal pressures, and confirmed alarm cleared."

Helpful Details to Include (If Known)

Final Thought

Continuous pressure alarms indicate a potentially dangerous condition. Always prioritize patient safety, rule out simple airflow restrictions first, and escalate appropriately when internal faults are suspected.

That is successful troubleshooting.