Drager Fabius GS

Alarm: EXP PORT LEAKAGE !!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

EXP PORT LEAKAGE !! alarm indicating unintended expiratory flow during inspiration, causing ventilation leak, inaccurate delivery, or inability to maintain airway pressure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Breathing Circuit for Leaks or Disconnections

Check Expiratory Valve Assembly

Check Expiratory Control Line (Hose Line)

Check Flow Sensor and Connections

Perform Flow Sensor Calibration

Re-test the System

If the Problem Persists

If the alarm continues after all external checks:

Clinical Use Tip

Never troubleshoot leak-related alarms on an active patient. Move the patient to a backup device before performing any troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit displaying EXP PORT LEAKAGE !! alarm with inability to maintain airway pressure during ventilation."

Cause

What was observed during troubleshooting.

Example:
"Leak traced to contaminated expiratory valve or no external issue found indicating possible internal failure."

Resolution

What action was taken.

Example:
"Cleaned and reseated expiratory valve and verified connections, or removed from service and sent for repair when issue persisted."

Helpful Details to Include (If Known)

Final Thought

Leak-related alarms should always be approached from external to internal causes. Proper inspection of circuits and valves combined with timely escalation ensures safe and effective troubleshooting.

That is successful troubleshooting.