Drager Fabius GS

Alarm: FLOW SENSOR FAIL !

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

FLOW SENSOR FAIL ! alarm indicating loss of flow measurement, resulting in inaccurate or absent ventilation monitoring and potential inability to safely deliver or verify patient breaths.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Flow Sensor Cable Connection

4. Inspect the Flow Sensor and Breathing System

5. Perform Flow Sensor Calibration

6. Substitute with a Known Good Flow Sensor

7. Inspect Environmental and Setup Factors

If the Problem Persists

If external causes have been ruled out and the alarm continues after cable checks, cleaning/replacement, and calibration:

Clinical Use Tip

Never troubleshoot or calibrate a flow sensor while connected to a patient. Always transition ventilation support before addressing this alarm.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit displayed FLOW SENSOR FAIL ! alarm with no tidal volume readings during setup."

Cause

What was observed during troubleshooting.

Example:
"Defective flow sensor identified after reseating cable and failed calibration."

Resolution

What action was taken.

Example:
"Replaced flow sensor, completed calibration successfully, and verified normal operation."

Helpful Details to Include (If Known)

Final Thought

Flow sensor alarms directly impact ventilation accuracy. Always rule out simple external causes like connections and contamination before escalating to internal failure, and prioritize patient safety throughout.

That is successful troubleshooting.