Drager Fabius GS

Alarm: FRESH GAS LOW !!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

FRESH GAS LOW !! alarm indicating insufficient fresh gas flow, which can lead to inadequate ventilation delivery and risk of patient hypoventilation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Medical Gas Supply Sources

Check Flowmeter and Gas Settings

Inspect Fresh Gas Hoses and Connections

Check Breathing Circuit for Leaks or Disconnections

Perform a Leak Test

Stop When Resolved

If the Problem Persists

If all external gas sources, hoses, and circuit issues have been ruled out and the alarm continues, an internal flow regulation or sensing failure is likely.

Clinical Use Tip

Never troubleshoot gas supply issues on an active patient. Always move the patient to a backup device before continuing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit displayed FRESH GAS LOW !! alarm with low delivered tidal volume during ventilation."

Cause

What was observed during troubleshooting.

Example:
"Kinked oxygen supply hose restricting fresh gas flow to the machine."

Resolution

What action was taken.

Example:
"Repositioned and replaced hose, verified normal gas flow and cleared alarm. Performed leak test and returned unit to service."

Helpful Details to Include (If Known)

Final Thought

Fresh gas flow is essential for safe anesthesia delivery. Always begin with supply and external causes before considering internal faults, and prioritize patient safety at every step.

That is successful troubleshooting.