Drager Fabius GS

Alarm: INSP PRES NOT REACH !!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

The INSP PRES NOT REACH !! alarm indicates that the inspiratory pressure is below the set PINSP value, which can lead to inadequate ventilation and failure to achieve target airway pressure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Fresh Gas Flow and Supply

Inspect Breathing Circuit for Leaks or Disconnections

Check APL Valve and Circuit Configuration

Evaluate Ventilator Settings

Check for Obstruction or Compliance Issues

Perform System Leak Test (if available)

If the Problem Persists

If all external causes have been ruled out and the alarm continues, this indicates a likely internal ventilator or pressure control issue.

Clinical Use Tip

Never troubleshoot pressure delivery issues on an active patient. Always move the patient to a backup device before investigating.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit alarming INSP PRES NOT REACH !! during ventilation; unable to achieve set inspiratory pressure."

Cause

What was observed during troubleshooting.

Example:
"Leak identified in breathing circuit connection causing loss of pressure."

Resolution

What action was taken.

Example:
"Reseated circuit connections and verified proper pressure delivery; alarm cleared."

Helpful Details to Include (If Known)

Final Thought

This alarm reflects a mismatch between expected and delivered pressure. Always rule out leaks and setup issues first, but escalate persistent failures as potential internal faults.

That is successful troubleshooting.