Drager Fabius GS

Alarm: O2 SUPPLY LOW !!!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

O2 SUPPLY LOW !!! alarm indicating oxygen supply pressure below minimum (~20 psi), risking inadequate oxygen delivery and unsafe ventilation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Wall Oxygen Supply

Inspect O2 Hose and Connections

Check Backup O2 Cylinder

Verify Cylinder Regulator Function

Isolate Supply Source

Stop When Resolved

If the Problem Persists

If all external sources (wall supply, hose, cylinder, regulator) are confirmed good and the alarm remains:

Clinical Use Tip

Never troubleshoot oxygen supply issues on an active patient. Move the patient to a backup device first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit alarming O2 SUPPLY LOW !!! with low displayed pipeline pressure and concern for inadequate oxygen delivery."

Cause

What was observed during troubleshooting.

Example:
"Kinked O2 supply hose restricting flow."

Resolution

What action was taken.

Example:
"Replaced damaged O2 hose; pressure restored and alarm cleared."

Helpful Details to Include (If Known)

Final Thought

Oxygen supply alarms must always be treated as critical. Start with simple supply checks, confirm true pressure loss, and escalate appropriately to protect patient safety and ensure reliable gas delivery.

That is successful troubleshooting.