Drager Fabius GS

Alarm: PEEP HIGH !

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

PEEP HIGH ! alarm indicating elevated airway pressure (>8 cmH₂O) in ManSpont mode, which can result in excessive airway pressure, impaired ventilation, and potential patient harm due to improper pressure release.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check APL Valve Setting

Check Fresh Gas Flow Rate

Check Breathing Circuit for Obstruction

Check APL Valve Function

Stop When Resolved

If the Problem Persists

If the alarm continues after external causes such as APL valve setting, fresh gas flow, and circuit obstruction have been ruled out:

Clinical Use Tip

Never troubleshoot pressure-related alarms while connected to a patient. Always transition the patient to a backup device before making adjustments.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit displaying PEEP HIGH ! alarm in ManSpont mode with elevated airway pressure and stiff reservoir bag."

Cause

What was observed during troubleshooting.

Example:
"APL valve improperly set causing excessive circuit pressure buildup."

Resolution

What action was taken.

Example:
"Adjusted APL valve to appropriate setting and verified normal airway pressure and alarm resolution. Unit returned to service."

Helpful Details to Include (If Known)

Final Thought

High PEEP alarms in manual or spontaneous modes are often caused by simple external factors like APL valve position or excessive gas flow. Always verify these first before escalating to internal issues, and prioritize patient safety at every step.

That is successful troubleshooting.