On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
POWER FAIL ! alarm indicating loss of mains power, causing the anesthesia machine to operate on battery and creating risk of complete device shutdown if power is not restored.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Immediately confirm the patient is being ventilated safely
- Switch to manual ventilation (bagging) or an alternate anesthesia machine if needed
- Ensure oxygen delivery is maintained
Verify the Reported Alarm and Behavior
- Confirm the POWER FAIL ! alarm is displayed
- Check if the unit is unplugged from wall power
- Observe if the machine is running on battery
- Monitor battery indicators for discharge
Check Power Cord Connection
- Ensure the AC power cord is firmly connected to the machine
- Ensure the cord is fully inserted into the wall outlet
- Reseat both ends of the connection
Verify Wall Outlet Functionality
- Plug in another device to confirm outlet power
- Move the machine to a known working outlet if needed
- Confirm consistent power at the outlet
Check Power Strip or Extension (If Used)
- Ensure power strip is turned ON
- Bypass the strip and plug directly into the wall
- Remove any unnecessary intermediaries
Inspect Power Cord Condition
- Check for cuts, fraying, or exposed wires
- Inspect for loose connectors or damaged plug
- Replace cord if damage is found
Confirm Internal Power Recognition
- Verify the machine indicates AC power is present
- Confirm the battery begins charging
- Ensure the alarm clears after power restoration
If the Problem Persists
If the POWER FAIL ! alarm continues despite confirmed external power:
- External causes have been ruled out
- This indicates a potential internal power supply or input failure
- Remove device from service
- Label Out of Service
- Send for repair or bench evaluation
Clinical Use Tip
Never troubleshoot on an active patient. Always move the patient to a backup device before addressing power-related issues.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"POWER FAIL ! alarm observed; unit running on battery with loss of mains power."
Cause
What was observed during troubleshooting.
Example:
"Power cord disconnected from wall outlet or outlet found to be non-functional."
Resolution
What action was taken.
Example:
"Reconnected power cord to verified working outlet; confirmed AC power restored, battery charging, and alarm cleared. Unit returned to service."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
Power-related alarms are often simple but critical. Always start with external checks, confirm power availability, and prioritize patient safety before escalating to internal failures.
That is successful troubleshooting.