Drager Fabius GS

Alarm: PRES THRESHOLD LOW !

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

PRES THRESHOLD LOW ! alarm indicating airway pressure alarm limits are set too low relative to ventilation settings, which can result in nuisance alarms or inaccurate pressure monitoring.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Current Ventilator Settings

Check Pressure Alarm Limits

Use AutoSet Function

Manually Adjust Alarm Limits (If Needed)

Confirm Alarm Resolution

If the Problem Persists

If the alarm continues after proper adjustment, external and configuration causes have been ruled out. Remove the device from service, label it Out of Service, and send it for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot alarm settings on an active patient. Always transition the patient to a backup device before making adjustments.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit displaying PRES THRESHOLD LOW ! alarm during ventilation after parameter changes."

Cause

What was observed during troubleshooting.

Example:
"Pressure alarm thresholds not aligned with updated ventilator settings."

Resolution

What action was taken.

Example:
"Used AutoSet to adjust alarm limits and verified proper operation. Alarm cleared."

Helpful Details to Include (If Known)

Final Thought

Alarm conditions like this are often due to configuration mismatches rather than equipment failure. Proper alignment of alarm thresholds ensures accurate monitoring and safe patient care.

That is successful troubleshooting.