Drager Fabius GS

Alarm: PRESSURE LIMITING !

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

PRESSURE LIMITING ! alarm indicating airway pressure reaching or exceeding PMAX, causing limited ventilation and potential risk of inadequate or unsafe pressure delivery.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Ventilator Pressure Settings (PMAX)

Check for Breathing Circuit Obstruction or Restriction

Check APL Valve and Fresh Gas Flow (if in Manual/Spont Mode)

Evaluate Patient or Test Lung Conditions

Check Ventilation Mode and Settings

Stop When Resolved

If the Problem Persists

External causes have been ruled out. Persistent pressure limiting may indicate internal pressure sensing or control issues.

Clinical Use Tip

Never troubleshoot on an active patient. Always transfer ventilation to a backup device before evaluating pressure-related alarms.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit displaying PRESSURE LIMITING ! alarm during ventilation with elevated airway pressures and limited breath delivery."

Cause

What was observed during troubleshooting.

Example:
"Breathing circuit obstruction or PMAX set too low causing pressure to reach limit; no internal faults if resolved externally."

Resolution

What action was taken.

Example:
"Adjusted PMAX and cleared circuit obstruction restoring normal operation, or removed unit from service for further evaluation."

Helpful Details to Include (If Known)

Final Thought

Pressure-related alarms must be addressed immediately. Always start with settings and external restrictions before escalating to internal issues.

That is successful troubleshooting.