Drager Fabius GS

Alarm: PRESSURE NEGATIVE !!!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

PRESSURE NEGATIVE !!! alarm indicating airway pressure ≤ -6.5 cmH₂O, often caused by circuit leak, disconnection, or improper ventilator settings.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check for Breathing Circuit Disconnection

4. Check for Leaks in the Breathing System

5. Check APL Valve and Fresh Gas Flow

6. Verify Ventilator Settings

7. Inspect Flow Sensor and Components

8. Isolate the Issue

9. Stop When Resolved

If the Problem Persists

If all external causes (circuit, leaks, settings, sensors) have been ruled out and the alarm continues:

Clinical Use Tip

Never troubleshoot a negative pressure condition on an active patient. Move the patient to a backup device before continuing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Unit displaying PRESSURE NEGATIVE !!! alarm with low/negative airway pressure during ventilation."

Cause

What was observed during troubleshooting.

Example:
"Leak identified in breathing circuit at Y-piece connection causing inability to maintain airway pressure."

Resolution

What action was taken.

Example:
"Reconnected and secured circuit connections, verified no leaks, confirmed normal airway pressure readings, and returned unit to service."

Helpful Details to Include (If Known)

Final Thought

Negative pressure alarms are often caused by simple issues like disconnections or leaks. Always start with external checks, verify system integrity, and escalate appropriately to ensure safe operation.

That is successful troubleshooting.