Drager Fabius GS

Alarm: RS232 COM1 FAIL !

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

RS232 COM1 FAIL ! alarm indicating loss of communication with an external monitor, resulting in missing data output or connectivity failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check External Monitor Connection

4. Inspect the RS232 Cable Condition

5. Verify External Monitor Power and Function

6. Check System Configuration Settings

7. Isolate the Issue

8. Stop When Resolved

If the Problem Persists

If all external connections, cables, and monitor functionality have been verified, the issue likely indicates internal COM1 port failure.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device before addressing communication issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"RS232 COM1 FAIL ! alarm present. External monitor not receiving data."

Cause

What was observed during troubleshooting.

Example:
"Loose RS232 cable connection at COM1 port."

Resolution

What action was taken.

Example:
"Reseated cable and verified secure connection. Alarm cleared and communication restored. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

Communication alarms often stem from simple connection issues but may indicate hardware failure. Always start with external checks, verify function, and escalate appropriately to maintain safe and reliable data flow.

That is successful troubleshooting.