Drager Fabius GS

Alarm: VENTILATOR FAIL !!!

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius GS

What This Guide Helps With

VENTILATOR FAIL alarm indicating ventilator mounting, sealing, or hose connection issue causing inability to deliver or control mechanical ventilation.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Ventilator Mounting and Seating

4. Inspect Ventilator Membrane and Seal

5. Check PEEP/PMAX Line Connection

6. Inspect Breathing System for Leaks or Disconnections

7. Reset Ventilation Mode

8. Stop When Resolved

If the Problem Persists

If all external connections, mounting, and seals have been verified, persistent VENTILATOR FAIL !!! alarm indicates likely internal ventilator module failure.

Remove device from service, label Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot ventilator-related alarms on an active patient. Always secure ventilation first before attempting any corrective actions.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"VENTILATOR FAIL !!! alarm present. Ventilator not initiating or maintaining mechanical ventilation."

Cause

What was observed during troubleshooting.

Example:
"Loose ventilator mounting / disconnected PEEP line or internal ventilator module failure indicated after external checks."

Resolution

What action was taken.

Example:
"Reseated ventilator module, secured membrane and cover, reconnected PEEP/PMAX line. If unresolved, removed from service and sent for repair."

Helpful Details to Include (If Known)

Final Thought

Ventilator failure alarms should always be taken seriously. Focus on sealing, connections, and mounting first, and escalate quickly if unresolved to protect patient safety.

That is successful troubleshooting.