On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Helps isolate oxygen-sensor condition, connection, calibration environment, gas-source, sampling, or configuration causes when oxygen readings are missing, incorrect, or cannot calibrate.
Step-by-Step Troubleshooting
1. Protect the Patient
Do not use an anesthesia machine for patient care when oxygen concentration cannot be monitored reliably. Provide an alternate verified anesthesia machine or oxygen-monitoring method before troubleshooting.
Expected outcome: Patient oxygen delivery and monitoring remain dependable while the Fabius is removed from active use.
2. Confirm the Exact Oxygen-Monitoring Problem
Determine whether the displayed oxygen concentration is absent, unstable, implausible, slow to respond, or whether calibration fails.
Compare the complaint with an independent approved oxygen analyzer when appropriate.
Expected outcome: The oxygen-monitoring problem is objectively confirmed.
3. Verify the Gas Environment
Confirm the machine is receiving the expected gas sources and that the breathing system is configured appropriately for the intended calibration or test.
Make sure unexpected fresh-gas mixtures or residual gases are not influencing the reading.
Expected outcome: The sensor is being evaluated in a valid and known gas environment.
4. Inspect the Oxygen Sensor and Mounting
Inspect the externally accessible oxygen sensor, sensor housing, adapter, and mounting location for looseness, damage, contamination, or improper seating.
Expected outcome: The sensor is properly installed and undamaged externally. If reseating restores a stable reading, proceed to calibration and final verification.
5. Inspect External Connections
Check any accessible cable, connector, sensor contact, or interface associated with the oxygen sensor for looseness, corrosion, strain, or visible damage.
Expected outcome: External sensor connections are intact and secure.
6. Verify Calibration Conditions
Follow the approved manufacturer calibration process under the proper gas conditions. Do not use undocumented service procedures or alter restricted calibration values.
Expected outcome: Calibration completes successfully. If so, compare the resulting reading with an independent analyzer.
7. Substitute a Known-Good Sensor When Appropriate
If compatible and permitted, install a known-good approved oxygen sensor and repeat the appropriate calibration and test.
Expected outcome: If the known-good sensor calibrates and reads correctly, replace the original sensor and proceed to final testing.
8. Compare Oxygen Reading With Independent Test Equipment
Use an approved gas analyzer to compare the displayed oxygen concentration under controlled conditions.
Expected outcome: The machine's indicated oxygen concentration tracks the independently measured concentration appropriately under the approved test procedure.
9. Verify Related Alarms and Functional Performance
After correction, verify oxygen monitoring over the required operating range according to approved testing and confirm applicable oxygen alarms and gas-delivery functions.
Expected outcome: Oxygen readings are stable, responsive, and consistent with test equipment. If so, troubleshooting is complete.
10. Escalate Persistent Calibration or Accuracy Problems
If correct installation, external connections, known gas conditions, and a known-good sensor do not resolve the problem, remove the machine from service.
Expected outcome: Further evaluation occurs at the qualified service level.
If the Problem Persists
Common external sensor, connection, calibration, gas-source, and test-condition causes have been ruled out. Remaining possibilities may involve internal oxygen-monitoring circuitry, signal processing, calibration functions, pneumatic sampling, or other service-level faults.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate manufacturer documentation and approved gas-analysis equipment
- Repaired or calibrated only by qualified personnel
Verify oxygen monitoring, gas delivery, alarms, ventilation, and the complete machine checkout before return to service.
Clinical Use Tip
Never assume the displayed oxygen concentration is correct after a sensor complaint; compare against an independent approved analyzer before return to clinical use.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported the oxygen monitor would not calibrate during the anesthesia-machine checkout."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the oxygen sensor was improperly seated at its external mounting point."
Resolution
What action was taken.
Example:
"The sensor was reseated, calibration completed successfully, and oxygen concentration was verified against approved gas-analysis equipment."
Helpful Details to Include (If Known)
- Displayed oxygen value
- Independent analyzer reading
- Calibration result
- Sensor age or condition if known
- Sensor seating
- Connector condition
- Gas source in use
- Known-good sensor result
- Alarm verification
- Final device status
- Displayed oxygen value
- Independent analyzer reading
- Calibration result
- Sensor age or condition if known
- Sensor seating
- Connector condition
- Gas source in use
- Known-good sensor result
- Alarm verification
- Final device status
Final Thought
Reliable oxygen monitoring is essential to anesthesia safety. Confirm the gas environment, sensor installation, connections, and calibration before suspecting internal electronics, and require objective verification before the machine returns to service.
That is successful troubleshooting.