Drager Fabius

PEEP Not Maintained or Pressure Drifts

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius

What This Guide Helps With

Helps identify external leaks, circuit problems, settings, accessories, or test conditions that can prevent stable PEEP or cause airway-pressure drift.

Step-by-Step Troubleshooting

1. Protect the Patient

Do not troubleshoot unstable PEEP or airway pressure while a patient relies on the affected ventilator. Move ventilation to an alternate verified source when pressure stability cannot be assured.

Expected outcome: Patient support is maintained and the affected unit is available for bench evaluation.

2. Confirm the Pressure Complaint

Determine whether PEEP falls below the selected value, rises unexpectedly, drifts over time, or differs from an external airway-pressure measurement.

Reproduce the condition using an appropriate test lung and analyzer.

Expected outcome: The pressure abnormality is confirmed under controlled conditions.

3. Verify the Selected Ventilation Settings

Confirm ventilation mode, PEEP setting, pressure settings, rate, and other externally accessible controls are configured as intended.

Check whether the complaint began after a mode or configuration change.

Expected outcome: The machine is operating with the intended settings. If an incorrect setting explains the complaint, correct it and proceed to verification.

4. Inspect the Breathing Circuit for Leaks

Examine the inspiratory and expiratory limbs, Y-piece, bag, filters, adapters, sampling connections, absorber interfaces, and other accessible breathing-system connections.

Expected outcome: No significant external leak is present. If a corrected leak restores stable PEEP, troubleshooting can stop after final testing.

5. Check for Restrictions or Obstructions

Inspect for kinked tubing, occluded filters, blocked expiratory components, water accumulation, or improperly installed accessories that could alter pressure.

Expected outcome: The breathing pathway is unobstructed and pressure is able to rise and fall normally.

6. Verify External Pressure Measurement

Use approved test equipment to compare the machine's indicated airway pressure with an independent measurement at an appropriate test point.

Expected outcome: Pressure indication and measured pressure are consistent with the applicable service evaluation. If the machine is stable and the discrepancy exists only in an external accessory, correct that accessory.

7. Compare With a Known-Good Circuit

When practical, replace the external breathing circuit and disposable accessories with compatible known-good components.

Expected outcome: If PEEP becomes stable, the removed circuit or accessory was contributing to the problem. Replace it and proceed to final verification.

8. Inspect Accessible Breathing-System Assemblies

Verify accessible valves, breathing-system assemblies, absorber components, and external sealing interfaces are correctly seated and undamaged.

Do not perform deep pneumatic disassembly.

Expected outcome: External assemblies are secure and properly installed.

9. Perform Final Functional Testing

Run the appropriate ventilator and breathing-system checks with a test lung. Confirm stable PEEP, expected pressure release, alarm behavior, and consistent ventilation over an adequate observation period.

Expected outcome: PEEP remains stable without unexplained drift. If so, the device may be returned to service after required checks.

10. Remove From Service if Pressure Remains Unstable

Persistent PEEP loss or pressure drift after external causes are eliminated requires escalation.

Expected outcome: The anesthesia machine is removed from service for qualified internal evaluation.

If the Problem Persists

External leaks, settings, restrictions, accessories, and test configuration have been ruled out. Remaining possibilities may involve internal pressure-control, expiratory-valve, pneumatic regulation, pressure-sensing, or calibration-related faults.

The device should be:

Verify ventilation, airway-pressure monitoring, alarms, and breathing-system integrity before return to service.

Clinical Use Tip

Unstable PEEP can compromise ventilation even when other displayed parameters appear normal; move the patient to verified support before investigating the machine.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"OR staff reported that PEEP would not remain stable during pre-case ventilation testing."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a damaged external circuit connection producing a breathing-system leak."

Resolution

What action was taken.

Example:
"The affected circuit component was replaced and stable PEEP, ventilation, alarms, and leak performance were verified before return to service."

Helpful Details to Include (If Known)

Final Thought

Pressure instability should be approached as a system problem first. Protect the patient, verify settings, eliminate leaks and restrictions, compare against approved test equipment, and escalate only after external causes are excluded and documented.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide