Drager Fabius

Touchscreen / Display Errors

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Fabius

What This Guide Helps With

This guide addresses situations where a Drager Fabius anesthesia machine has touchscreen or display issues such as:

The focus is on simple, external, and easily verifiable causes first—such as power quality, peripheral connections, environmental factors, and user interaction—before assuming internal display or control board failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify AC Power and Battery Status

Confirm the unit is plugged directly into a known-good outlet.

Check for tripped breakers or recently used red emergency outlets that may have unstable supply.

Review battery status on the screen (if visible).

Why this matters: Low voltage or unstable power can cause screen freezes, flickering, or boot issues.

3. Perform a Controlled Reboot

Observe:

If the screen restores after reboot, the issue may have been a software hang.

4. Inspect for External Physical Causes

Touchscreens can become unresponsive due to fluid intrusion or contamination.

5. Check for Peripheral Lockups

If physical controls work but touchscreen does not, the issue may be isolated to the display module.

6. Evaluate Environmental Factors

Overheating or condensation can cause temporary display malfunction.

7. Observe for Backlight vs. Display Failure

This helps differentiate between backlight failure and total display loss.

If the Problem Persists

If power has been verified, reboot performed, external causes ruled out, and the touchscreen or display remains unresponsive or unstable:

Likely causes at this point include:

Do not attempt internal disassembly beyond standard Clinical Engineering access panels.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a touchscreen issue on an actively anesthetized patient.

If the display is unreliable:

An unreliable display means unreliable monitoring and ventilator control feedback.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Touchscreen not responding to touch in OR 3. Screen frozen during case setup.”

Cause

What was observed during troubleshooting.

Example:
“Unit powered but touchscreen unresponsive. Reboot temporarily restored function. Issue recurred after 15 minutes. No external contamination noted.”

Resolution

What action was taken.

Example:
“Removed anesthesia machine from service. Labeled Out of Service. Scheduled vendor repair for suspected display module failure.”

Helpful Details to Include

Final Thought

Display issues on an anesthesia machine are not minor inconveniences—they directly impact safe ventilation and monitoring. Always approach troubleshooting logically: verify power, rule out external factors, reboot safely, and escalate appropriately. Strong documentation protects both the patient and the department.

That is successful troubleshooting.