Drager Infinity Delta Series

Battery Not Charging or Power / UPS Failure

On this page

Asset Type

Patient Monitor

Manufacturer

Drager

Model

Infinity Delta Series

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where the Infinity Delta monitor battery is not charging, the unit does not stay powered when unplugged, or there are power/UPS-related failures. Symptoms may include low battery alarms, failure to charge, unexpected shutdowns, or no power-on condition. The focus is on verifying external power sources, connections, and battery conditions before assuming internal failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify AC Power Source

Check Power Cord and Connections

Confirm Power Indicator Lights

Check Docking Station or Mount (if applicable)

Inspect Battery Condition

Perform a Power Cycle

Test with a Known-Good Battery

Check for UPS / Internal Power Alarms

Verify Operation on AC Only

If the Problem Persists

If all external power sources, cables, docking connections, and battery checks have been verified, the issue is likely internal (battery charging circuit, internal power supply, or UPS components).

At this point:

Recognizing when the issue is beyond external troubleshooting is proper Clinical Engineering practice.

Clinical Use Tip

Do not troubleshoot power issues on a monitor connected to an active patient. Always transfer the patient to a properly functioning monitor before removing power or swapping batteries to avoid loss of monitoring.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Monitor not charging battery and shuts down when unplugged from AC power.”

Cause

What was observed during troubleshooting.

Example:
“Battery failed to accept charge and did not hold power; verified with known-good battery test.”

Resolution

What action was taken.

Example:
“Replaced defective battery and confirmed proper charging and operation on battery power.”

Helpful Details to Include (If Known)

Final Thought

Power-related issues are often caused by simple external factors like cords, outlets, or worn batteries. A structured approach ensures these are ruled out before escalating. Prioritizing patient safety, minimizing downtime, and clearly documenting findings are essential parts of effective Clinical Engineering troubleshooting.

That is successful troubleshooting.