Drager Infinity Delta Series

Network or central station connectivity loss

On this page

Asset Type

Patient Monitor

Manufacturer

Drager

Model

Infinity Delta Series

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where the Infinity Delta monitor loses connection to the network or central station. Symptoms may include loss of patient visibility at the central station, intermittent disconnects, network alarms, or no communication with charting systems. The focus is on verifying external connections, network infrastructure, and configuration before assuming internal failure.

Step-by-Step Troubleshooting

Verify the Monitor is Powered On and Fully Booted

Check Network Cable Connection (Hardwired Systems)

Inspect the Network Cable for Damage

Verify Wall Port Functionality

Confirm Correct Network Jack is Being Used

Check Network Settings on the Monitor

Restart the Monitor

Check Central Station Status

Look for Network or Communication Alarms

Verify Device is Assigned/Admitted Correctly

Check for Recent Network Changes

If the Problem Persists

If all external connections, cables, ports, and network conditions have been verified, the issue is likely internal to the monitor (network interface hardware or internal communication board).

Knowing when to stop prevents unnecessary risk and wasted troubleshooting time.

Clinical Use Tip

Do not troubleshoot network connectivity while the monitor is actively in use on a patient. Move the patient to a properly functioning monitor first to maintain continuous central monitoring and patient safety.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
Monitor not showing on central station; staff reports intermittent disconnects.

Cause

What was observed during troubleshooting.

Example:
Loose Ethernet cable connection identified at monitor; no link light observed.

Resolution

What action was taken.

Example:
Reseated and replaced Ethernet cable; verified network connection restored and monitor visible on central station.

Helpful Details to Include

Final Thought

Effective network troubleshooting starts with simple physical checks before escalating to configuration or internal faults. Maintaining patient visibility at the central station is critical, so always prioritize safe patient transfer before troubleshooting. Clear documentation ensures repeat issues can be identified and resolved faster.

That is successful troubleshooting.