On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Flow sensor errors, failed calibration, or inaccurate ventilation readings caused by contamination, improper assembly, or sensor-related external issues.
Step-by-Step Troubleshooting
Ensure Patient Safety First
Remove the ventilator from patient use and transition the patient to a functioning ventilator before troubleshooting.
Expected: Patient ventilation continues safely on an alternate device.
Why it matters: Flow measurement errors can result in incorrect tidal volumes and unsafe ventilation.
Verify the Reported Issue
Power on the ventilator and observe the exact error message or calibration failure.
- Specific flow sensor or calibration error message
- Whether error occurs during startup or operation
Expected: Clear identification of the error condition.
Why it matters: Confirms whether the issue is repeatable and sensor-related.
Inspect Flow Sensor Assembly
Locate and inspect the flow sensor and associated components.
- Ensure the flow sensor is properly seated and connected
- Check for visible contamination (moisture, secretions, debris)
- Confirm correct orientation per manufacturer design
Expected: Clean, dry, and properly installed sensor.
Why it matters: Misalignment or contamination is a common cause of sensor errors.
Check Patient Circuit and Accessories
Inspect the breathing circuit connected to the ventilator.
- Verify correct circuit setup (no missing or incorrect components)
- Check for condensation or water buildup in tubing
- Replace HME/filter if saturated
- Ensure no kinks or obstructions
Expected: Dry, unobstructed circuit with proper components.
Why it matters: Excess moisture or restriction affects flow readings and calibration.
Perform Flow Sensor Cleaning (If Applicable)
If contamination is present, clean the flow sensor per hospital protocol.
- Remove the sensor
- Clean and dry thoroughly (no residual moisture)
- Reinstall securely
Expected: Sensor is clean and completely dry before reuse.
Why it matters: Even small amounts of moisture can disrupt flow measurement accuracy.
Repeat Calibration Procedure
Initiate the ventilator’s calibration or self-test process.
- Ensure no patient is connected
- Follow on-screen calibration prompts
- Allow full completion without interruption
Expected: Calibration completes successfully with no errors.
Why it matters: Confirms whether the issue was due to external or setup-related causes.
Swap with Known Good Flow Sensor (If Available)
If the issue persists, replace the flow sensor with a known good unit.
Expected: Device calibrates successfully with replacement sensor.
Why it matters: Helps isolate whether the original sensor is faulty.
Check Connector and Port Condition
Inspect the connection point where the flow sensor interfaces with the ventilator.
- Look for bent pins, corrosion, or debris
- Ensure secure and proper connection
Expected: Clean, intact connection interface.
Why it matters: Poor electrical or mechanical connection can trigger sensor errors.
If the Problem Persists
If all external causes (sensor condition, circuit setup, moisture, and connections) have been ruled out, the issue is likely internal (sensor interface circuitry or processing fault).
Remove the ventilator from service, label it Out of Service, and send it for repair or bench evaluation.
Knowing when to stop prevents unnecessary risk and is part of proper troubleshooting.
Clinical Use Tip
Never attempt calibration or troubleshooting with a patient connected. Always move the patient to a confirmed working ventilator first.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Ventilator displaying flow sensor error and failing calibration during startup."
Cause
What was observed during troubleshooting.
Example:
"Contaminated flow sensor with moisture present affecting calibration."
Resolution
What action was taken.
Example:
"Cleaned and dried flow sensor, reinstalled, and successfully completed calibration test."
Helpful Details to Include (If Known)
- Calibration pass/fail result
- Presence of moisture or debris in sensor
- Circuit condition (wet/dry, filter status)
- Sensor swapped (yes/no)
- Error message text
- Final device status (returned to service or sent for repair)
Final Thought
Flow sensor issues are often caused by simple external factors like moisture or improper setup. A structured, step-by-step approach ensures patient safety while avoiding unnecessary repairs. Proper escalation and clear documentation are essential to effective Clinical Engineering practice.
That is successful troubleshooting.