Drager Oxylog 3000 Plus

Pre-Use Check or Leak Test Failure

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Oxylog 3000 Plus

What This Guide Helps With

Pre-use check or leak test failing due to circuit leaks, loose connections, incorrect setup, or accessory issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient is supported on a safe, functioning device.

Why it matters: Leak issues can compromise ventilation accuracy and patient safety.

Verify the Reported Failure

Expected: Clear identification of the failure step.

Why it matters: Helps isolate whether the issue is circuit-related or device-related.

Inspect the Patient Circuit Assembly

Expected: Circuit is intact with no visible leaks.

Why it matters: Most leak failures are caused by circuit integrity issues.

Check the Test Lung and Accessories

Expected: Proper closed system for testing.

Why it matters: An incomplete or open system will always fail a leak test.

Inspect Filters, Water Traps, and Valves

Expected: All components properly installed with no gaps.

Why it matters: Misaligned or loose accessories introduce leaks.

Check Inspiratory and Expiratory Ports

Expected: Tight, secure connections at ventilator ports.

Why it matters: Even small leaks at ports can cause test failure.

Replace the Circuit with a Known Good Setup

Expected: Leak test passes with known good circuit.

Why it matters: Confirms whether issue is circuit vs ventilator.

Check for Environmental or Setup Errors

Expected: Proper configuration for testing.

Why it matters: Incorrect setup can mimic a leak condition.

If the Problem Persists

If all external components, circuits, and setup conditions have been verified and the leak test still fails, the issue is likely internal (e.g., internal valves, pressure sensors, or pneumatic system).

Remove the ventilator from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop prevents unnecessary risk and ensures patient safety.

Clinical Use Tip

Never attempt to troubleshoot or run leak tests while the ventilator is connected to a patient. Always use a test lung and remove from service first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator failed pre-use check with leak test error."

Cause

What was observed during troubleshooting.

Example:
"Leak identified in patient circuit due to loose expiratory limb connection."

Resolution

What action was taken.

Example:
"Secured connections and replaced circuit; pre-use check passed successfully."

Helpful Details to Include (If Known)

Final Thought

Pre-use check failures are most commonly caused by external setup issues, not internal failures. A methodical approach—starting with the circuit and accessories—ensures safe, efficient troubleshooting. Recognizing when the issue is internal and escalating appropriately protects both the patient and the integrity of the device.

That is successful troubleshooting.