Drager Perseus

Battery or UPS Not Charging or Short Runtime

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Perseus

What This Guide Helps With

Troubleshooting failed battery charging, unexpectedly short backup runtime, loss of mains recognition, external UPS problems, or battery-related alarms.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot an unresolved power or battery problem while the Perseus is supporting an active patient.

If power reliability becomes questionable during a procedure:

Expected: Patient ventilation and anesthesia delivery are not dependent on equipment with uncertain backup power.

Why it matters: A Perseus battery failure may result in loss of ventilation if the machine shuts down after mains power is removed.

If the patient has been safely transferred or the machine is not in clinical use, continue troubleshooting.

2. Confirm the Exact Failure Mode

Determine whether the reported problem involves:

Record the exact alarm wording, charge percentage, operating state, and whether the power cord was connected.

Expected: The complaint is separated into a mains-input, internal-battery, external-UPS, or accessory-power issue.

3. Verify the Wall Receptacle

Test the receptacle with an approved electrical tester or known-good biomedical test load.

Confirm:

Move the Perseus to a verified compatible receptacle when appropriate.

Expected: The receptacle supplies stable power and the machine recognizes mains operation.

If charging begins or the mains indication returns, the external power source was the likely cause. Stop troubleshooting and correct or report the facility electrical issue.

4. Inspect and Reseat the Power Cord

Disconnect the machine from clinical use before manipulating the power connection.

Inspect the power cord for:

Reconnect the cord securely at the machine and wall outlet. Confirm it is routed through the intended strain-relief path and is not connected through an added power strip.

Expected: The power cord remains firmly seated and the mains indicator appears on the status display.

Drager specifies that the internal batteries charge automatically from mains power and directs users to verify both mains and battery indications after connecting the power cable.

If reseating or replacing a damaged cord restores charging, complete electrical-safety and functional testing before returning the machine to service.

5. Check the Mains and Battery Status Indications

Power the machine normally and observe the status display.

Verify:

A “Battery low” alarm indicates less than 20% charge when mains power is unavailable. A “Battery very low” condition indicates less than 10% charge or may accompany a disconnected battery or failed battery fuse. Internal battery or fuse conditions require service evaluation rather than external troubleshooting.

Expected: Stable mains recognition is present and the battery charge indication responds normally.

If mains power is verified externally but not recognized by the machine, remove the unit from service and escalate for repair.

6. Check the Operating Temperature

Determine whether the machine was:

Move the machine to an approved room-temperature environment and allow it to stabilize while connected to verified mains power.

Expected: The temperature alarm clears and normal charging resumes.

Drager states that charging may be suspended when battery temperature is high and that prolonged operation above 35 °C may prevent proper charging or reduce available battery power.

If the alarm remains after the device reaches normal room temperature, escalate for bench evaluation.

7. Allow a Complete Charging Period

For a depleted battery, a machine returning from storage, or an uncertain charge history:

Do not assume the battery has failed based on a brief charging period.

Expected: The indicated charge level increases and reaches an acceptable level without battery alarms.

Drager specifies at least eight hours to reach full battery power. Battery capacity also decreases with age and use.

If the charge level does not increase after verified mains connection and an adequate charging period, remove the machine from service.

8. Evaluate an External UPS Separately

When the complaint refers to a room UPS, isolation panel, ceiling supply, or emergency-power system, determine whether that system is separate from the Perseus internal battery.

Check:

Do not open or internally service a facility UPS unless that work is within Clinical Engineering’s authorization and training.

Expected: The external UPS supplies stable output or the problem is referred to Facilities, Electrical Services, or the responsible vendor.

If the external UPS is operating normally but the Perseus still does not recognize mains power, continue with machine-specific evaluation.

9. Perform an Approved Battery Functional Test

After a complete charge, perform only the battery verification permitted by the current facility procedure and applicable Drager service documentation.

Before removing mains power:

A new, fully charged battery is specified for a minimum backup time of 30 minutes, with a typical value of 150 minutes; actual runtime may vary with battery condition and operating load.

Expected: The machine transfers to battery without interruption and demonstrates acceptable performance under the approved test method.

If the machine shuts down unexpectedly, charge drops rapidly, or runtime fails the approved acceptance criterion, stop testing and remove the device from service.

10. Verify Battery Recall or Correction Status

Review the machine’s serial number, service history, recall records, and battery-replacement documentation.

The FDA identified a Class I correction involving Perseus internal backup batteries that could fail unexpectedly during battery operation without a preceding low-battery alarm. The listed customer actions included testing device batteries and arranging replacement for machines that failed the test.

Expected: Documentation confirms that all applicable battery corrections and required testing have been completed.

If correction status cannot be confirmed, contact Drager or the authorized service provider before returning the machine to clinical use.

11. Review Battery Age and Service History

Check the CMMS and service records for:

Do not access or replace internal batteries without the required authorization, training, service documentation, and post-repair testing.

Expected: Battery history supports continued use, or an aged or repeatedly failing battery is identified for authorized replacement.

If an approved battery replacement resolves the issue, complete all required operational, alarm, battery-transfer, and electrical-safety testing before return to service.

If the Problem Persists

Verified mains power, the power cord, ambient temperature, charging time, external UPS operation, and operating conditions have been checked.

The remaining cause is likely an internal battery, battery fuse, charging circuit, power-input assembly, temperature-sensing circuit, or other internal power-system failure.

The device should be:

Do not continue repeated unplug testing on a machine that has demonstrated unexpected shutdown or unreliable battery operation. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never test battery runtime during an active anesthetic. Transfer the patient to verified equipment first and keep manual ventilation immediately available during any power-related event.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Drager Perseus battery remained at a low charge and the machine shut down shortly after mains power was disconnected."

Cause

What was observed during troubleshooting.

Example:
"Verified mains power was present and the power cord was secure, but the internal battery failed to gain charge after eight hours and failed the approved battery functional test."

Resolution

What action was taken.

Example:
"Removed the Perseus from service, labeled it Out of Service, provided a replacement anesthesia machine, and referred the unit for authorized battery and charging-system evaluation."

Helpful Details to Include (If Known)

Final Thought

Power troubleshooting on an anesthesia machine must prioritize uninterrupted ventilation and safe anesthesia delivery. Verify external power first, confirm adequate charging conditions, test backup operation only in a controlled setting, and escalate unreliable internal power systems promptly. Clear CCR documentation preserves the evidence needed for repair and patient-safety follow-up.

That is successful troubleshooting.