Drager Perseus

Breathing System Leak or Leak Test Failure

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Perseus

What This Guide Helps With

Troubleshooting failed leakage tests or excessive breathing-system leakage caused by loose circuits, absorbers, sampling components, vaporizers, seals, or incorrect assembly.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform a system test, leakage test, or extended troubleshooting while a patient is connected to the anesthesia machine.

If a leak is suspected during a procedure:

Expected: Patient ventilation and anesthetic delivery are maintained using verified equipment.

The Perseus pressurizes the breathing system during testing, and the manufacturer specifically warns not to perform the system test with a patient connected.

2. Confirm the Exact Failure

Review the test screen and determine whether the reported problem is:

Record any displayed test details, affected subsystem, leakage value, and color-coded result.

Expected: The complaint is narrowed to a repeatable test condition rather than a general report that the machine “leaks.”

3. Confirm the Device Is Prepared Exactly as Directed

Place the Perseus in Standby and prepare it according to the on-screen checklist.

Verify:

Repeat the applicable automatic test.

Expected: The test completes successfully when the machine is configured correctly.

If the test passes, document the configuration error and stop troubleshooting.

4. Inspect the Breathing Circuit and Patient Connections

Check the inspiratory and expiratory hoses, Y-piece, filters, adapters, test plug, and breathing bag.

Look for:

Reseat each connection firmly. Replace questionable disposable components with verified compatible items, then repeat the leakage test.

Expected: Leakage decreases and the test passes after a defective or poorly seated circuit component is corrected.

5. Check the CO2 Absorber Assembly

Inspect the CO2 absorber or CLIC absorber assembly.

Verify:

The manufacturer identifies an improperly secured absorber or CLIC adapter as a possible leakage source.

Repeat the leakage test after reseating or replacing the absorber assembly.

Expected: The test passes when the absorber and its sealing surfaces are correctly seated.

6. Inspect the APL Valve and Breathing-Bag Connection

Verify that the APL valve is correctly installed, securely seated, and positioned as required by the on-screen test instructions.

Inspect the breathing-bag connection, bag elbow, and optional flexible bag arm.

Check for:

Do not disassemble the valve internally. Correct only external fitment or replace a visibly damaged approved component.

Expected: The leak test passes after the APL valve or bag assembly is correctly seated.

7. Check the Water Trap and Sample Line

Inspect the gas-monitoring sampling path.

Verify:

Replace the sample line or water trap with verified compatible components when damage is suspected.

Expected: The leakage test passes after the sampling system is restored to a sealed condition.

8. Inspect External Breathing-System Seating

If the failure began after cleaning or breathing-system replacement, verify that the removable breathing system is fully seated and locked into the machine.

Check externally that:

Do not proceed into internal pneumatic disassembly.

Expected: The test passes after correcting an incomplete installation or externally visible seal problem.

9. Isolate Vaporizers as a Possible Leak Source

Ensure each vaporizer is:

Run the leakage test with the vaporizers configured as directed by the on-screen instructions. When appropriate and permitted by facility procedure, test one vaporizer at a time or remove a suspect vaporizer to determine whether the failure follows it.

The Perseus leakage assistant can support evaluation of breathing-system, circuit, and vaporizer leakage by continuously displaying test pressure and leakage changes.

Expected: The failure is isolated to a particular vaporizer or mounting position.

If the test passes without one vaporizer, remove that vaporizer from service for separate evaluation.

10. Use the Leakage Assistant

From the test details screen, open the Leakage Assistant and follow the displayed instructions.

While the system maintains test pressure:

The manufacturer recommends systematic leak localization by isolating individual components and using the leakage assistant when needed.

Expected: A specific circuit, absorber, sampling component, vaporizer, or breathing-system connection is identified.

If the test passes, complete a full operational checkout before returning the machine to service.

If the Problem Persists

If the breathing circuit, bag, absorber, sampling components, vaporizer configuration, accessory connections, and external breathing-system seating have been verified, common external causes have been ruled out.

The remaining condition may involve an internal valve, pneumatic seal, flow-sensor seal, breathing-system housing defect, mounting interface, or another internal leakage source.

The device should be:

Do not repeatedly accept a failed or restricted test result to keep the machine in service. Knowing when to stop external troubleshooting and escalate is proper Clinical Engineering practice.

Clinical Use Tip

Never troubleshoot a suspected anesthesia breathing-system leak on an active patient. Transfer ventilation and anesthetic delivery first, then test the affected machine in a controlled environment.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Drager Perseus failed its pre-use leakage test and displayed excessive breathing-system leakage."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the CO2 absorber incompletely seated, preventing the absorber seal from closing properly."

Resolution

What action was taken.

Example:
"Reseated and locked the absorber, repeated the leakage and system tests successfully, completed an operational checkout, and returned the machine to service."

Helpful Details to Include (If Known)

Final Thought

Breathing-system leak failures require a methodical approach because loose external components can imitate serious internal faults. Protect the patient first, isolate one component at a time, and escalate when the machine cannot pass its required testing. Accurate CCR documentation preserves both safety and service history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide