Drager Perseus

CO2 or Anesthetic Agent Gas Measurement Unavailable

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Perseus

What This Guide Helps With

Troubleshooting unavailable CO2 or anesthetic-agent measurements caused by sample-line, filter, water-trap, startup-test, accessory, or internal gas-sensor faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended gas-measurement troubleshooting while the Perseus is supporting an active patient.

If CO2 or anesthetic-agent monitoring becomes unavailable during a procedure:

Drager specifies that O2, CO2, and anesthetic gases must be monitored whenever the breathing system is used. If the internal patient-gas measurement module fails, an independent gas-measurement system is required.

Expected: Patient care is no longer dependent on the unavailable internal gas measurements.

If alternate monitoring or equipment resolves the immediate safety concern, continue troubleshooting off patient.

2. Confirm the Exact Failure Mode

Review what clinical staff observed and record the complete alarm message.

Determine whether:

The Perseus identifies disconnected or blocked sampling components separately from internal gas-sensor failures.

Expected: The complaint is narrowed to a sampling-path problem, isolated measurement problem, or complete gas-analyzer failure.

If the reported condition was resolved after correcting an obvious disconnected sampling line, verify operation and stop.

3. Confirm Normal Power and Startup Status

With the machine removed from active patient use:

Do not interrupt power while the machine is supporting a patient.

Expected: The machine starts normally and either passes the gas-measurement portion of the system test or produces a repeatable failure.

If gas measurement returns and the complete test passes, perform a controlled functional verification before returning the unit to service.

4. Check the Sample-Line Connections

Inspect both ends of the sampling line.

Verify that the line is:

Drager directs that the sample line be securely connected at the patient-side sampling point and water trap, with no adapters used. An improperly connected or damaged line can cause faulty measurements.

Expected: The sample line is firmly connected and correctly routed.

If reseating the line restores stable CO2 and agent measurements, complete the system test and stop.

5. Inspect the Sample Line for Occlusion or Damage

Examine the full length of the sampling line for:

Replace a questionable line with an approved new sample line rather than attempting to flush, clean, or disinfect it. Drager identifies sample lines as single-use items and warns that disinfectants can damage the sampling components.

Expected: A clean, dry, unobstructed sampling path is established.

If replacing the sample line restores the measurements, verify operation and stop.

6. Check the Patient-Side Filter or Sampling Port

Inspect the filter, HME, elbow, or adapter where the sample line connects.

Look for:

Replace a wet, blocked, contaminated, or questionable patient-side accessory with a compatible component.

The manufacturer’s remedies for sample-line disconnection or occlusion specifically include checking the sample line, water trap, and patient-side filter.

Expected: Gas can move freely from the patient connection through the sampling line.

If the values return after replacing the filter or HME, complete verification and stop.

7. Inspect the Water Trap

Check whether the water trap is:

Empty or replace the water trap according to its instructions for use. Reseat it fully and confirm that any Water trap disconnected? or Water trap full alarm clears.

A full, damaged, contaminated, or improperly connected water trap can interrupt or distort gas measurement.

Do not apply silicone spray, aerosol cleaner, or lubricant to the water-trap holder or O-rings. Drager warns that silicone can disrupt measurement and silicone or aerosol residue may create a fire risk.

Expected: The water trap is clean, correctly seated, within its replacement interval, and free of excess fluid.

If replacement restores measurements, complete the system test and stop.

8. Distinguish Agent Measurement From Vaporizer Recognition

If CO2 measurement works but anesthetic-agent information is unavailable, determine whether the complaint involves:

Inspect the vaporizer externally and confirm that it is:

The optional Vapor View system provides vaporizer identification, dial-position, filling-level, and prediction information. A Vapor View failure is therefore different from a failure of the patient-gas measurement module.

Expected: The complaint is correctly classified as vaporizer recognition, agent delivery, or measured-agent concentration failure.

If the problem is limited to a persistent Vapor View hardware alarm, document it separately and escalate according to the displayed alarm guidance.

9. Perform the System Test

After correcting or replacing external sampling components:

Drager warns that a device with nonoperational CO2 or anesthetic-agent sensors requires suitable substitute monitoring before clinical use.

Expected: The system test completes successfully without unresolved gas-measurement faults.

If the system test repeatedly reports a sensor failure, stop troubleshooting and escalate.

10. Perform a Controlled Measurement Verification

When authorized test equipment is available, verify gas-measurement performance using a certified gas source or analyzer test setup according to facility procedures.

Confirm:

Do not use exhaled breath, improvised chemicals, or uncontrolled anesthetic-agent delivery as a test source.

The manufacturer’s safety-check specifications use certified test-gas concentrations to verify CO2 and anesthetic-agent accuracy and also require sampling-flow and leakage evaluation of the patient-gas measurement module.

Expected: Gas values are present, stable, and reasonable for the approved test source.

If verification passes and all required safety testing is complete, return the machine to service according to facility policy.

If the Problem Persists

If the sample line, patient-side filter, water trap, vaporizer condition, power state, and system-test setup have been checked, common external causes have been ruled out.

A persistent Gas sensor failure, repeated CO2 sensor accuracy low, failed system-test result, or inability to detect approved test gas suggests a fault involving the internal patient-gas measurement module or internal sampling pathway.

The device should be:

Do not open the gas-measurement module or attempt board-level repair without appropriate authorization, training, documentation, and test equipment.

Knowing when external troubleshooting is complete and escalation is required is proper troubleshooting.

Clinical Use Tip

Never troubleshoot missing CO2 or anesthetic-agent monitoring while relying on the affected machine for an active patient.

Move the patient to another verified anesthesia machine or establish approved independent gas monitoring before continuing. Anesthesia delivery without dependable ventilation and agent monitoring presents a serious patient-safety risk.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"OR staff reported that the Perseus displayed no CO2 waveform or sevoflurane concentration during pre-use setup."

Cause

What was observed during troubleshooting.

Example:
"The water trap was full and the sample line contained condensate, preventing the patient-gas measurement module from obtaining a sample."

Resolution

What action was taken.

Example:
"Replaced the water trap and sample line, completed the system test, verified CO2 and anesthetic-agent measurement with an approved test setup, and returned the machine to service."

Helpful Details to Include (If Known)

Final Thought

Unavailable gas measurements require a safety-first response and a logical inspection of the complete sampling path. Checking connections, consumables, moisture, test results, and vaporizer conditions can resolve many external problems without unnecessary disassembly. Persistent sensor failures require removal from service, qualified repair, and complete CCR documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide