Drager Perseus

Flow Sensor Calibration Failure or Unstable Volume Measurement

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Perseus

What This Guide Helps With

Troubleshooting failed flow-sensor calibration or unstable tidal and minute-volume readings caused by test status, leaks, moisture, sensor seating, contamination, or sensor failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the Perseus is connected to an active patient.

Coordinate with the anesthesia provider to:

Expected: Patient care is no longer dependent on equipment with questionable volume measurement.

Why it matters: Failed flow sensing may affect ventilation and monitoring. The system test must not be performed while a patient is connected because the breathing system is pressurized during testing.

If safe backup ventilation and monitoring are established, continue.

2. Confirm the Exact Complaint

Review the reported message and determine whether the problem involves:

Record whether the problem began after breathing-system removal, reprocessing, flow-sensor replacement, or circuit changes.

Expected: The affected sensor and operating condition are identified.

A disconnected or replaced breathing system, an uncalibrated sensor, and flow-sensor failure may cause inspiratory or expiratory flow-sensor calibration alarms.

3. Check Calibration and System-Test Status

From Standby, review the most recent system-test results and determine whether flow-sensor calibration is overdue or incomplete.

The Flow sensor calibration required message may occur when:

Run the guided breathing-system test or full system test as appropriate.

Expected: The test completes successfully and the calibration message clears.

The breathing-system test calibrates the flow sensors, includes leakage testing, and should be completed after flow-sensor or breathing-system replacement.

If calibration passes and measurements stabilize, document the result and stop.

4. Inspect the External Breathing Circuit

Check the complete test configuration for:

Reconnect or replace questionable external accessories and repeat the leakage or breathing-system test.

Expected: The circuit passes its leakage check and volume readings remain stable over repeated breaths.

Why it matters: Circuit leakage can reduce the delivered or measured volume and create differences between inspired and expired readings.

If correcting a circuit problem resolves the measurement instability, stop.

5. Check Flow-Sensor Installation

With the machine in Standby and prepared according to facility procedures, inspect the accessible inspiratory and expiratory flow-sensor assemblies.

Verify that:

Do not force the sensors or ports.

Expected: Both sensors are correctly installed without looseness, missing seals, or visible misalignment.

Improperly installed or damaged flow sensors and missing O-rings may cause breathing-system leakage.

After reseating a sensor, repeat the breathing-system test. If the test passes, stop.

6. Inspect for Moisture, Deposits, or Damage

Visually inspect each accessible flow sensor for:

Do not insert tools, brushes, compressed air, or test probes into the sensor.

Confirm that any recently reprocessed component is completely dry and was processed according to its specific instructions.

Expected: Sensors are clean, dry, and visibly undamaged.

Deposits or reprocessing contamination may damage the measuring elements. Replace the sensor when deposits remain or visible damage is present.

7. Substitute a Known-Good Flow Sensor

When permitted by facility policy, replace the suspected inspiratory or expiratory sensor with a compatible, known-good sensor.

Replace one sensor at a time when practical so the failed component can be identified.

After each replacement:

Expected: Calibration succeeds and stable volume measurement returns after replacing the defective sensor.

If replacement resolves the issue, stop and retain the failed sensor according to facility procedures.

8. Verify Volume Performance

Connect a calibrated anesthesia-machine analyzer and approved test lung.

Using a controlled test configuration:

Expected: Measurements remain repeatable and within approved limits without flow-sensor alarms.

Do not return the machine to service based only on a cleared alarm. Confirm actual functional performance.

9. Review Test Results and Device Logs

Record:

If the test identifies another correctable external issue, address it and repeat the complete test.

Expected: The final device status and failure pattern are clearly established.

If the Problem Persists

If the calibration repeatedly fails or volume measurement remains unstable after circuit verification, correct sensor installation, successful leak correction, and known-good sensor substitution, common external causes have been ruled out.

The problem may involve internal sensor connections, breathing-system components, pressure measurement, pneumatic control, or internal electronics.

The device should be:

Do not continue replacing accessories without evidence or attempt internal pneumatic or board-level repair outside approved service procedures.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never evaluate unstable volume measurements on an active patient. Transfer the patient to verified equipment first and perform calibration, leakage, and analyzer testing only in a controlled environment.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported repeated “Expiratory flow sensor not calibrated” alarms and unstable expired tidal-volume readings during pre-use testing."

Cause

What was observed during troubleshooting.

Example:
"Expiratory flow sensor had visible residue and continued to fail the breathing-system test after reseating; the circuit passed leakage testing."

Resolution

What action was taken.

Example:
"Replaced the expiratory flow sensor, completed a successful breathing-system test, and verified stable tidal and minute-volume measurements with a calibrated analyzer."

Helpful Details to Include (If Known)

Final Thought

Reliable anesthesia volume measurement depends on correctly installed, clean flow sensors and a leak-free breathing system. Patient safety comes first, followed by controlled testing, logical accessory substitution, performance verification, and timely escalation when external causes have been eliminated.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide