On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting failed flow-sensor calibration or unstable tidal and minute-volume readings caused by test status, leaks, moisture, sensor seating, contamination, or sensor failure.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
Do not perform extended troubleshooting while the Perseus is connected to an active patient.
Coordinate with the anesthesia provider to:
- Transfer ventilation to another verified anesthesia machine or approved backup method.
- Maintain appropriate independent patient monitoring.
- Move the affected Perseus to Standby before testing.
Expected: Patient care is no longer dependent on equipment with questionable volume measurement.
Why it matters: Failed flow sensing may affect ventilation and monitoring. The system test must not be performed while a patient is connected because the breathing system is pressurized during testing.
If safe backup ventilation and monitoring are established, continue.
2. Confirm the Exact Complaint
Review the reported message and determine whether the problem involves:
- Flow sensor calibration required
- Inspiratory flow sensor not calibrated
- Expiratory flow sensor not calibrated
- Unstable inspiratory tidal volume
- Unstable expiratory tidal volume
- Large or inconsistent differences between inspired and expired volume
- Fluctuating minute-volume readings
- Volume-controlled ventilation failure
Record whether the problem began after breathing-system removal, reprocessing, flow-sensor replacement, or circuit changes.
Expected: The affected sensor and operating condition are identified.
A disconnected or replaced breathing system, an uncalibrated sensor, and flow-sensor failure may cause inspiratory or expiratory flow-sensor calibration alarms.
3. Check Calibration and System-Test Status
From Standby, review the most recent system-test results and determine whether flow-sensor calibration is overdue or incomplete.
The Flow sensor calibration required message may occur when:
- The sensors have not been calibrated since power-up.
- More than 24 hours have passed since the last calibration.
- The breathing system or sensor was removed after calibration.
Run the guided breathing-system test or full system test as appropriate.
Expected: The test completes successfully and the calibration message clears.
The breathing-system test calibrates the flow sensors, includes leakage testing, and should be completed after flow-sensor or breathing-system replacement.
If calibration passes and measurements stabilize, document the result and stop.
4. Inspect the External Breathing Circuit
Check the complete test configuration for:
- Loose inspiratory or expiratory hose connections
- Cracked, stretched, or incorrectly connected hoses
- Loose Y-piece, test lung, filter, or breathing bag
- Incorrectly seated absorber
- Open sampling ports
- Excessive condensate
- Occluded or saturated filters
- Incorrect circuit configuration for the selected patient category
Reconnect or replace questionable external accessories and repeat the leakage or breathing-system test.
Expected: The circuit passes its leakage check and volume readings remain stable over repeated breaths.
Why it matters: Circuit leakage can reduce the delivered or measured volume and create differences between inspired and expired readings.
If correcting a circuit problem resolves the measurement instability, stop.
5. Check Flow-Sensor Installation
With the machine in Standby and prepared according to facility procedures, inspect the accessible inspiratory and expiratory flow-sensor assemblies.
Verify that:
- Each sensor is installed in the correct port.
- Both ports are fully inserted and tightened.
- The sensor body is not reversed, tilted, or partially seated.
- Required seals and O-rings are present.
- The surrounding breathing-system housing is seated correctly.
Do not force the sensors or ports.
Expected: Both sensors are correctly installed without looseness, missing seals, or visible misalignment.
Improperly installed or damaged flow sensors and missing O-rings may cause breathing-system leakage.
After reseating a sensor, repeat the breathing-system test. If the test passes, stop.
6. Inspect for Moisture, Deposits, or Damage
Visually inspect each accessible flow sensor for:
- Condensate
- Disinfectant residue
- Lint or fibers
- Dried secretions
- Aerosol deposits
- Cracks, discoloration, or deformation
- Damaged internal measuring elements
Do not insert tools, brushes, compressed air, or test probes into the sensor.
Confirm that any recently reprocessed component is completely dry and was processed according to its specific instructions.
Expected: Sensors are clean, dry, and visibly undamaged.
Deposits or reprocessing contamination may damage the measuring elements. Replace the sensor when deposits remain or visible damage is present.
7. Substitute a Known-Good Flow Sensor
When permitted by facility policy, replace the suspected inspiratory or expiratory sensor with a compatible, known-good sensor.
Replace one sensor at a time when practical so the failed component can be identified.
After each replacement:
- Confirm correct installation.
- Run the breathing-system test.
- Review the sensor-specific test result.
- Recheck displayed volume stability.
Expected: Calibration succeeds and stable volume measurement returns after replacing the defective sensor.
If replacement resolves the issue, stop and retain the failed sensor according to facility procedures.
8. Verify Volume Performance
Connect a calibrated anesthesia-machine analyzer and approved test lung.
Using a controlled test configuration:
- Deliver several consistent breaths.
- Observe inspired tidal volume, expired tidal volume, and minute volume.
- Watch for drifting or breath-to-breath instability.
- Check whether the problem occurs in multiple ventilation modes.
- Compare displayed and analyzer measurements using applicable manufacturer specifications and facility acceptance limits.
Expected: Measurements remain repeatable and within approved limits without flow-sensor alarms.
Do not return the machine to service based only on a cleared alarm. Confirm actual functional performance.
9. Review Test Results and Device Logs
Record:
- Exact alarm text
- Inspiratory or expiratory sensor involved
- System-test and breathing-system-test results
- Leakage result
- Sensors or accessories exchanged
- Analyzer readings
- Whether the failure followed reprocessing or breathing-system removal
If the test identifies another correctable external issue, address it and repeat the complete test.
Expected: The final device status and failure pattern are clearly established.
If the Problem Persists
If the calibration repeatedly fails or volume measurement remains unstable after circuit verification, correct sensor installation, successful leak correction, and known-good sensor substitution, common external causes have been ruled out.
The problem may involve internal sensor connections, breathing-system components, pressure measurement, pneumatic control, or internal electronics.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for qualified bench evaluation or Drager-authorized repair
Do not continue replacing accessories without evidence or attempt internal pneumatic or board-level repair outside approved service procedures.
Knowing when to stop is proper troubleshooting.
Clinical Use Tip
Never evaluate unstable volume measurements on an active patient. Transfer the patient to verified equipment first and perform calibration, leakage, and analyzer testing only in a controlled environment.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Anesthesia staff reported repeated “Expiratory flow sensor not calibrated” alarms and unstable expired tidal-volume readings during pre-use testing."
Cause
What was observed during troubleshooting.
Example:
"Expiratory flow sensor had visible residue and continued to fail the breathing-system test after reseating; the circuit passed leakage testing."
Resolution
What action was taken.
Example:
"Replaced the expiratory flow sensor, completed a successful breathing-system test, and verified stable tidal and minute-volume measurements with a calibrated analyzer."
Helpful Details to Include (If Known)
- Exact alarm or message displayed
- Inspiratory or expiratory sensor affected
- Last successful system-test date
- Breathing system recently removed or reprocessed
- Flow sensor recently replaced
- Sensor seating and O-ring condition
- Circuit and leakage-test results
- Presence of condensate or deposits
- Known-good sensor substitution result
- Analyzer and test-lung results
- Final device status
- Repair or service escalation reference
Final Thought
Reliable anesthesia volume measurement depends on correctly installed, clean flow sensors and a leak-free breathing system. Patient safety comes first, followed by controlled testing, logical accessory substitution, performance verification, and timely escalation when external causes have been eliminated.
That is successful troubleshooting.