Drager Perseus

Fresh Gas Flow Control Failure

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Perseus

What This Guide Helps With

Troubleshooting unresponsive, inaccurate, unstable, or absent fresh-gas flow caused by supply, control, configuration, circuit, or gas-mixer issues.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the Perseus is supporting an active patient.

If fresh-gas delivery becomes unreliable during a case:

Expected: Patient care is transferred to dependable equipment before Clinical Engineering evaluation begins.

Why it matters: On electronically controlled configurations, ventilation may continue after fresh-gas delivery fails, but the gas mixer may provide no fresh gas. Emergency oxygen delivery or a manual resuscitator may be required.

If patient safety and backup equipment are established, continue troubleshooting off patient.

2. Confirm the Exact Failure Mode

Ask staff what occurred and record the exact alarm or symptom.

Determine whether the complaint involved:

Expected: The complaint is separated into a gas-supply, control-input, measurement, leakage, or internal-delivery problem.

3. Identify the Installed Gas-Mixer Configuration

Determine whether the unit uses:

Do not apply mechanical-control troubleshooting to an electronic mixer or vice versa.

Expected: Testing is matched to the actual configuration.

4. Verify Medical Gas Supply Connections

Inspect the O₂, Air, and optional N₂O pipeline hoses.

Confirm that:

Test the wall outlet and hose with approved equipment when a supply-pressure problem is suspected.

Expected: All required gases show stable, acceptable supply pressure.

If reseating or replacing a defective hose restores normal operation, complete the system test and stop troubleshooting.

5. Check Electrical Power and Device Status

For electronically controlled units:

Do not assume a gas-mixer failure when the device is operating from a depleted battery or has experienced an incomplete startup.

Expected: The device has stable electrical power and normal control-interface operation.

If restoring mains power resolves the control problem, allow the battery to charge and complete the required system test before returning the machine to service.

6. Inspect Fresh-Gas Control Settings

For an electronically controlled mixer:

For a mechanically controlled mixer:

Do not force a stuck valve or continue using a cracked, loose, or damaged control knob.

Expected: The control input changes smoothly and the indicated fresh-gas flow responds appropriately.

7. Check the Emergency and Auxiliary Oxygen Controls

On electronically controlled configurations, inspect the O₂ switch and flowmeter.

Confirm that:

Expected: Routine fresh-gas control is not being confused with separate emergency or auxiliary oxygen flow.

If correcting an inadvertently active Add. O₂ control restores normal operation, complete a system test and controlled functional verification before returning the unit to service.

8. Inspect the Breathing Circuit and Breathing Bag

Check for conditions that can resemble a fresh-gas control failure:

Correct visible problems and repeat the system test.

Expected: The breathing system fills normally and does not generate a fresh-gas-low or leakage alarm.

9. Check the Vaporizer Installation

Inspect the installed vaporizer without opening or servicing it.

Confirm that:

If a vaporizer concern exists, remove it from use according to facility procedure and repeat testing with an approved known-good vaporizer or bypass configuration.

Expected: The vaporizer is secure and is not interfering with fresh-gas delivery.

Stop immediately if liquid anesthetic leakage, strong odor, physical damage, or an abnormal control condition is found.

10. Perform the Manufacturer System Test

Place the machine in the required standby or test condition and perform the normal system test using the current Drager instructions for use.

Before starting:

Record whether the fresh-gas delivery, gas supply, leakage, and mixer portions pass or fail.

Expected: The complete system test passes without fresh-gas delivery alarms.

A fresh-gas delivery failure may remain active until the system test is successfully completed. A persistent failure after testing requires qualified service evaluation.

11. Verify Fresh-Gas Output in a Controlled Test

Only after the system test passes:

Expected: Set flow, displayed flow, and independently measured output agree within the applicable manufacturer specification.

Do not return the device to service based only on screen indications when the original complaint involved inaccurate delivery.

12. Reproduce the Original Complaint

Repeat the operating condition reported by staff without connecting a patient.

Confirm that:

Expected: The original problem cannot be reproduced and all required tests pass.

If the issue is resolved, document the cause, test results, and final device status, then stop troubleshooting.

If the Problem Persists

If gas sources, hoses, electrical power, control settings, emergency oxygen controls, breathing-circuit components, vaporizer installation, and system-test preparation have been verified, common external causes have been ruled out.

The likely problem may involve the internal:

Do not force controls, enter unauthorized service modes, adjust internal regulators, or disassemble the gas-mixing system.

The anesthesia machine should be:

A recurring fresh-gas delivery failure or failed system test is sufficient reason to keep the machine out of clinical service. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot unstable fresh-gas delivery while the machine remains responsible for an active patient. Transfer care first and preserve independent oxygen, ventilation, anesthetic-gas, and physiological monitoring.

Emergency oxygen controls are a temporary clinical contingency, not proof that the normal fresh-gas delivery system is safe for continued use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Perseus fresh-gas flow setting would not respond and oxygen continued flowing after the electronic flow was reduced."

Cause

What was observed during troubleshooting.

Example:
"The emergency O₂ switch was found in the Add. O₂ position with the separate O₂ flow-control valve open, making the normal electronic fresh-gas controls appear unresponsive."

Resolution

What action was taken.

Example:
"Returned the emergency O₂ controls to the normal position, completed the system test, verified fresh-gas flow and oxygen concentration with approved test equipment, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

Fresh-gas flow problems require immediate attention because displayed ventilation may continue even when normal gas delivery is compromised. Protect the patient first, verify external supplies and controls logically, independently test performance, and escalate persistent mixer or valve failures. Thorough CCR documentation supports safe return to service and effective repair.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide