Drager Perseus

High Airway Pressure or Continuous Pressure Alarm

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Perseus

What This Guide Helps With

Troubleshooting elevated airway-pressure alarms caused by blocked circuits, restrictive accessories, incorrect settings, APL valve position, scavenging problems, or equipment faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the Perseus is ventilating an active patient.

Have the anesthesia provider:

Expected: Patient ventilation is maintained without depending on equipment displaying unresolved pressure alarms.

Why it matters: High airway pressure can result from either a patient condition or an equipment or circuit problem. Clinical causes must be evaluated by the anesthesia provider before Clinical Engineering begins equipment troubleshooting.

If alternate ventilation is established, continue evaluation away from the patient.

2. Confirm the Exact Alarm and Operating Conditions

Record the complete alarm message before clearing or restarting the machine.

Determine whether the display shows:

Also document:

Expected: The specific alarm condition and circumstances are clearly identified.

3. Inspect the Patient Circuit for Obstruction

With the machine removed from patient use, inspect the complete breathing pathway.

Check for:

Straighten or replace any questionable component.

Expected: The inspiratory and expiratory pathways are open and unrestricted.

If removing the obstruction resolves the alarm during controlled testing, complete the system test and stop troubleshooting.

4. Check Filters, HMEFs, and Added Accessories

Inspect all filters, HMEFs, adapters, sampling connectors, and other accessories installed in the circuit.

Look for:

Replace suspect disposable components with approved known-good accessories.

Expected: The circuit operates normally without unnecessary resistance.

If replacing a restrictive accessory resolves the alarm, complete functional testing and stop.

5. Verify Circuit Connections and Component Seating

Confirm that:

Do not perform internal disassembly or force components into position.

Expected: The breathing system and circuit are assembled correctly with no visible obstruction or misconnection.

6. Inspect the Breathing Bag and Bag Hose

Check that the breathing bag:

Expected: The breathing bag moves freely without restricting the system.

If repositioning or replacing the bag assembly resolves the issue, complete the required system test and stop.

7. Verify Ventilation Settings and the Paw Alarm Limit

Have the anesthesia provider confirm the clinical settings when the event occurred. During bench evaluation, use facility-approved test settings.

Review:

Confirm that the upper Paw alarm limit is appropriate for the expected test pressure and has not been accidentally set below the normal PIP.

Do not independently change prescribed patient settings merely to suppress an alarm.

Expected: The alarm limit is appropriate, and the selected ventilation settings do not intentionally produce excessive pressure.

8. Check the APL Valve in MAN/SPON Mode

The APL valve affects pressure limitation during manual ventilation and spontaneous breathing.

When evaluating MAN/SPON operation:

Expected: Pressure is properly relieved and limited during an approved manual-ventilation test.

If the APL valve binds, does not relieve pressure, or behaves inconsistently, remove the machine from service.

9. Check the O2 Flush Control

Inspect the O2 flush button for:

During controlled testing, verify that the button releases immediately when no longer pressed.

Expected: The O2 flush operates only while intentionally activated and returns normally.

If the control sticks or activates unexpectedly, stop testing and remove the machine from service.

10. Inspect the Anesthetic Gas Scavenging Path

Check the external scavenging system for:

Compare the setup with a known-good operating room connection when available. Do not adjust internal scavenging components without the applicable service procedure and authorization.

Expected: Waste gas can leave the machine without an external restriction.

If correcting the external scavenging connection resolves the alarm, complete functional testing and stop.

11. Test With Known-Good Accessories

Install an approved known-good:

Use controlled adult or pediatric test settings appropriate for the test equipment.

Observe:

Expected: Normal pressure delivery without repeated pressure alarms.

If the machine passes with known-good accessories, the original circuit or accessory configuration was likely responsible.

12. Perform the System Test and Final Functional Check

After replacing or reconnecting breathing-system components:

Expected: The Perseus completes its system test and ventilates normally without unexplained elevated pressure.

Return the machine to service only after all required tests pass.

If the Problem Persists

If the alarm continues with known-good circuits, filters, accessories, test lungs, appropriate settings, and an unrestricted scavenging connection, common external causes have been ruled out.

A fault may involve the:

The machine should be:

Do not perform unauthorized internal disassembly. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a persistent high-pressure alarm while the machine remains responsible for an active patient. Establish alternate ventilation first and allow the anesthesia provider to assess patient, airway, and tube-related causes.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported repeated high airway pressure and continuously high airway pressure alarms during pre-use ventilation testing."

Cause

What was observed during troubleshooting.

Example:
"Inspection found a wet, occluded expiratory filter that restricted circuit flow and caused elevated airway pressure during test-lung ventilation."

Resolution

What action was taken.

Example:
"Replaced the filter, inspected the breathing circuit and scavenging connections, completed the Perseus system test, and verified normal ventilation and pressure-alarm operation with a test lung."

Helpful Details to Include (If Known)

Final Thought

High airway pressure may originate from the patient, breathing circuit, settings, scavenging pathway, or anesthesia machine. Protect the patient first, eliminate visible restrictions logically, verify operation with known-good accessories, and escalate persistent faults with complete documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide