Drager Perseus

O2 Sensor Calibration Failure or Incorrect FiO2 Reading

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Perseus

What This Guide Helps With

Troubleshooting failed O2-sensor calibration or implausible FiO2 values caused by gas supply, sampling-path, water-trap, circuit, test, or internal measurement faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended troubleshooting while the anesthesia machine is connected to an active patient.

If the displayed FiO2 is questionable during a procedure:

Expected: Patient ventilation and oxygen monitoring are maintained using verified equipment.

If patient care has been safely transferred, continue troubleshooting off patient.

2. Confirm the Exact Failure Being Reported

Review the alarm message, staff report, system-test result, and device logbook.

Determine whether the complaint involves:

Expected: The issue is classified as a gas-supply, fresh-gas delivery, sampling-path, or patient-gas measurement problem.

These alarms represent different conditions. An O2 sensor failure may involve the patient-gas measurement system, while O2 supply and fresh-gas flow alarms involve separate delivery systems.

3. Verify Power and Allow a Complete Startup

Connect the machine to a known-good AC receptacle and confirm normal mains-power indication.

Restart the device only when permitted by facility procedure. Allow it to complete its startup sequence without interruption.

Check for:

Expected: The machine reaches Standby normally and is ready to perform its system test.

If a normal restart clears the condition and subsequent testing passes, complete functional verification, document the result, and stop.

4. Check the Oxygen Supply

Inspect the external O2 supply before assuming that the sensor has failed.

Verify:

Expected: A stable medical O2 source is available to the machine without supply-related alarms.

An absent or unstable O2 supply can cause unexpected oxygen values and failed system testing without indicating an internal O2 sensor defect.

If correcting the supply restores normal readings and testing, document the cause and stop.

5. Inspect the Sample Line

Follow the sample line from the patient connection to the water trap.

Check for:

Confirm the line is securely attached at the approved sampling location and at the water trap.

Expected: The sample line is open, dry, correctly routed, and securely connected.

A damaged, blocked, or improperly connected sample line can produce faulty gas measurements.

Replace a questionable disposable sample line. If readings normalize, complete testing and stop.

6. Inspect the Water Trap

Check the water trap for:

Empty or replace the water trap according to the applicable instructions. Ensure the replacement is fully seated.

Do not apply silicone spray or unapproved substances to the water-trap holder or seals.

Expected: The water trap is clean, undamaged, correctly seated, and below its maximum fill level.

A full, contaminated, damaged, or improperly seated water trap can interfere with patient-gas measurement.

If replacement corrects the FiO2 reading, complete the system test and stop.

7. Check the Breathing Circuit and Fresh-Gas Settings

Verify that:

Expected: The breathing circuit is correctly assembled and produces a stable, plausible oxygen concentration.

At low fresh-gas flows, the measured inspiratory oxygen concentration may not immediately equal the selected fresh-gas concentration. Allow the circuit concentration to stabilize before declaring the reading inaccurate.

If correcting the circuit or allowing adequate stabilization resolves the discrepancy, document and stop.

8. Check for Contamination or Recent Reprocessing

Ask whether the problem started after:

Inspect only externally accessible components. Do not open the patient-gas measurement module.

Expected: No evidence is found of liquid, aerosol, disinfectant, silicone, or other contamination entering the sampling path.

Evidence of internal contamination requires repair evaluation.

9. Perform the Full System Test

Place the machine in Standby with no patient connected.

Prepare the machine according to the on-screen checklist and initiate the complete system test. Do not substitute only a breathing-system leakage test when the complaint involves O2 calibration.

If the test stops:

Expected: The O2 measurement and associated gas-measurement functions display a successful result.

If the complete test passes, continue with independent verification before returning the machine to service.

10. Verify the FiO2 Reading Independently

Using the facility’s approved performance-verification procedure:

Expected: The displayed FiO2 agrees with the independent analyzer within the applicable tolerance.

If the machine passes consistently, return it to service according to department policy.

11. Check for Intermittent Failure

Repeat the system test or controlled gas verification sufficiently to determine whether the problem is intermittent.

Review:

Expected: The problem either remains corrected or can be reproduced reliably.

Multiple inaccurate gas values may indicate a broader patient-gas measurement problem rather than an isolated O2 issue.

If the Problem Persists

If the O2 calibration continues to fail, the FiO2 value remains outside tolerance, or an O2 sensor failure returns after the gas supply, sample line, water trap, circuit, and system test have been checked, the patient-gas measurement module or associated internal components may be faulty.

The device should be:

Do not attempt internal sensor, pneumatic, or board-level repairs without the required training, documentation, and test equipment. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot an inaccurate FiO2 reading while the machine remains clinically relied upon. Move the patient to verified equipment or establish independent oxygen monitoring before Clinical Engineering begins evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia reported that the Perseus displayed an unexpectedly low FiO2 and failed the O2 portion of the system test."

Cause

What was observed during troubleshooting.

Example:
"The water trap was overfilled and the sample line contained moisture, restricting gas sampling and producing an inaccurate oxygen measurement."

Resolution

What action was taken.

Example:
"Removed the machine from service, replaced the water trap and sample line, completed the full system test, verified FiO2 against a calibrated analyzer, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

Incorrect oxygen readings require a safety-first, system-based approach. Verify the supply, sampling components, circuit, settings, and system-test results before suspecting the internal measurement module. Appropriate escalation and clear CCR documentation protect both the patient and the next technician.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide