On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting failed O2-sensor calibration or implausible FiO2 values caused by gas supply, sampling-path, water-trap, circuit, test, or internal measurement faults.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
Do not perform extended troubleshooting while the anesthesia machine is connected to an active patient.
If the displayed FiO2 is questionable during a procedure:
- Notify the anesthesia provider immediately.
- Use an independent oxygen-monitoring system.
- Transfer ventilation and anesthesia delivery to another verified machine when directed.
- Do not rely on a questionable FiO2 value for clinical decisions.
Expected: Patient ventilation and oxygen monitoring are maintained using verified equipment.
If patient care has been safely transferred, continue troubleshooting off patient.
2. Confirm the Exact Failure Being Reported
Review the alarm message, staff report, system-test result, and device logbook.
Determine whether the complaint involves:
- O2 sensor failure
- FiO2 low or FiO2 high
- O2 fresh-gas flow measurement failure
- No O2 delivery
- O2 supply low
- An O2-related system-test or calibration failure
- A displayed FiO2 that does not match an independent analyzer
Expected: The issue is classified as a gas-supply, fresh-gas delivery, sampling-path, or patient-gas measurement problem.
These alarms represent different conditions. An O2 sensor failure may involve the patient-gas measurement system, while O2 supply and fresh-gas flow alarms involve separate delivery systems.
3. Verify Power and Allow a Complete Startup
Connect the machine to a known-good AC receptacle and confirm normal mains-power indication.
Restart the device only when permitted by facility procedure. Allow it to complete its startup sequence without interruption.
Check for:
- Power interruptions during startup
- Abnormal boot messages
- Additional sensor or gas-measurement alarms
- A system test that was canceled or bypassed
Expected: The machine reaches Standby normally and is ready to perform its system test.
If a normal restart clears the condition and subsequent testing passes, complete functional verification, document the result, and stop.
4. Check the Oxygen Supply
Inspect the external O2 supply before assuming that the sensor has failed.
Verify:
- The O2 pipeline hose is fully connected at the wall outlet and machine inlet.
- The pipeline outlet provides the expected supply pressure.
- The backup O2 cylinder is properly mounted and sufficiently filled.
- Cylinder valves and pressure indications agree with the machine status display.
- No O2 supply or delivery alarm is present.
Expected: A stable medical O2 source is available to the machine without supply-related alarms.
An absent or unstable O2 supply can cause unexpected oxygen values and failed system testing without indicating an internal O2 sensor defect.
If correcting the supply restores normal readings and testing, document the cause and stop.
5. Inspect the Sample Line
Follow the sample line from the patient connection to the water trap.
Check for:
- Loose connections
- Cracks or cuts
- Kinks or compression
- Moisture accumulation
- Internal blockage
- Incorrect routing
- Nonapproved adapters or sample tubing
- Connection to an inappropriate circuit port
Confirm the line is securely attached at the approved sampling location and at the water trap.
Expected: The sample line is open, dry, correctly routed, and securely connected.
A damaged, blocked, or improperly connected sample line can produce faulty gas measurements.
Replace a questionable disposable sample line. If readings normalize, complete testing and stop.
6. Inspect the Water Trap
Check the water trap for:
- Overfilling
- Condensate
- Blood or secretion contamination
- Cracks or physical damage
- Loose seating
- Damaged seals
- Incorrect or incompatible water-trap type
Empty or replace the water trap according to the applicable instructions. Ensure the replacement is fully seated.
Do not apply silicone spray or unapproved substances to the water-trap holder or seals.
Expected: The water trap is clean, undamaged, correctly seated, and below its maximum fill level.
A full, contaminated, damaged, or improperly seated water trap can interfere with patient-gas measurement.
If replacement corrects the FiO2 reading, complete the system test and stop.
7. Check the Breathing Circuit and Fresh-Gas Settings
Verify that:
- Inspiratory and expiratory hoses are connected to the correct ports.
- The breathing circuit has no obvious disconnection or major leak.
- The CO2 absorber is properly installed.
- The fresh-gas O2 setting matches the intended bench-test condition.
- There is sufficient fresh-gas flow to change the circuit concentration.
- The circuit has been allowed enough time to reach the newly selected concentration.
Expected: The breathing circuit is correctly assembled and produces a stable, plausible oxygen concentration.
At low fresh-gas flows, the measured inspiratory oxygen concentration may not immediately equal the selected fresh-gas concentration. Allow the circuit concentration to stabilize before declaring the reading inaccurate.
If correcting the circuit or allowing adequate stabilization resolves the discrepancy, document and stop.
8. Check for Contamination or Recent Reprocessing
Ask whether the problem started after:
- Water-trap replacement
- Breathing-system reassembly
- Cleaning or disinfection
- Aerosol or nebulizer use
- Liquid entering the sample line
- A major condensation event
Inspect only externally accessible components. Do not open the patient-gas measurement module.
Expected: No evidence is found of liquid, aerosol, disinfectant, silicone, or other contamination entering the sampling path.
Evidence of internal contamination requires repair evaluation.
9. Perform the Full System Test
Place the machine in Standby with no patient connected.
Prepare the machine according to the on-screen checklist and initiate the complete system test. Do not substitute only a breathing-system leakage test when the complaint involves O2 calibration.
If the test stops:
- Record the displayed component and failure message.
- Correct any identified external condition.
- Repeat the affected test.
- Do not accept an O2-related irregularity merely to obtain a completed test result.
Expected: The O2 measurement and associated gas-measurement functions display a successful result.
If the complete test passes, continue with independent verification before returning the machine to service.
10. Verify the FiO2 Reading Independently
Using the facility’s approved performance-verification procedure:
- Connect a test lung or approved test setup.
- Compare the Perseus FiO2 reading with a calibrated independent oxygen analyzer.
- Verify performance at ambient air and at a controlled high-O2 concentration.
- Allow both systems to stabilize before comparing values.
- Apply the acceptance limits in the current manufacturer service documentation.
Expected: The displayed FiO2 agrees with the independent analyzer within the applicable tolerance.
If the machine passes consistently, return it to service according to department policy.
11. Check for Intermittent Failure
Repeat the system test or controlled gas verification sufficiently to determine whether the problem is intermittent.
Review:
- Previous system-test results
- Alarm history
- Whether the reading drifts over time
- Whether movement of the sample line affects the value
- Whether the problem returns after warm-up
- Whether multiple gases are reading incorrectly
Expected: The problem either remains corrected or can be reproduced reliably.
Multiple inaccurate gas values may indicate a broader patient-gas measurement problem rather than an isolated O2 issue.
If the Problem Persists
If the O2 calibration continues to fail, the FiO2 value remains outside tolerance, or an O2 sensor failure returns after the gas supply, sample line, water trap, circuit, and system test have been checked, the patient-gas measurement module or associated internal components may be faulty.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for authorized repair or bench evaluation
- Evaluated using the current Drager service procedure
- Returned to service only after successful system testing and independent O2 verification
Do not attempt internal sensor, pneumatic, or board-level repairs without the required training, documentation, and test equipment. Knowing when to stop is proper troubleshooting.
Clinical Use Tip
Never troubleshoot an inaccurate FiO2 reading while the machine remains clinically relied upon. Move the patient to verified equipment or establish independent oxygen monitoring before Clinical Engineering begins evaluation.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Anesthesia reported that the Perseus displayed an unexpectedly low FiO2 and failed the O2 portion of the system test."
Cause
What was observed during troubleshooting.
Example:
"The water trap was overfilled and the sample line contained moisture, restricting gas sampling and producing an inaccurate oxygen measurement."
Resolution
What action was taken.
Example:
"Removed the machine from service, replaced the water trap and sample line, completed the full system test, verified FiO2 against a calibrated analyzer, and returned the unit to service."
Helpful Details to Include (If Known)
- Exact alarm or system-test message
- Whether the problem occurred during startup or patient use
- O2 pipeline and cylinder pressure status
- Selected fresh-gas O2 concentration and flow
- Displayed FiO2 value
- Independent analyzer value
- Sample-line condition and replacement status
- Water-trap condition and fill level
- Evidence of moisture or contamination
- Complete system-test result
- Whether the problem was intermittent
- Software version
- Final device status
Final Thought
Incorrect oxygen readings require a safety-first, system-based approach. Verify the supply, sampling components, circuit, settings, and system-test results before suspecting the internal measurement module. Appropriate escalation and clear CCR documentation protect both the patient and the next technician.
That is successful troubleshooting.