Drager Perseus

Scavenging / AGSS Restriction or Evacuation Fault

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Perseus

What This Guide Helps With

Troubleshooting weak, excessive, or restricted anesthetic-gas evacuation caused by hose routing, connections, AGSS terminal suction, accessories, or receiving-system faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not perform extended scavenging troubleshooting while the Perseus is supporting an active patient.

If evacuation is inadequate, anesthetic odor is present, or the scavenging flow indicator is outside its acceptable range:

Expected: Patient care continues on functioning equipment, and personnel are not unnecessarily exposed to anesthetic gas.

If another device is placed into service, continue troubleshooting only in a controlled Clinical Engineering setting.

2. Confirm the Exact Failure Condition

Determine what was reported or observed:

Record whether the problem is constant or occurs only when the machine is moved.

Expected: The complaint is clearly identified before components are replaced.

3. Verify the AGS and Wall-Terminal Connections

Confirm that:

An incorrectly connected disposal system can allow anesthetic gas to enter the surrounding area.

Expected: Both ends remain securely connected and evacuation flow is restored.

If the indicator returns to its normal range, complete the system checkout and stop troubleshooting.

4. Inspect the Scavenging Hose and Routing

Examine the complete hose for:

Reposition the machine and route the hose without restricting its diameter.

Expected: The hose remains open and unrestricted throughout its entire length.

If correcting the routing restores normal flow, secure the hose properly and stop.

5. Test With a Known-Good Approved Hose

When available, substitute a compatible, known-good scavenging hose.

Do not use tubing or adapters that have not been approved for the machine and facility AGSS configuration.

Expected:

6. Observe the Perseus AGS Flow Indicator

Activate the facility scavenging system according to the approved checkout process and observe the AGS flow indicator.

During normal operation and system testing, the indicator should remain within the marked normal range.

Interpret abnormal positions as follows:

Expected: The flow indicator remains stable within the normal marked range.

Do not adjust a medical-gas pipeline terminal or ejector unless authorized and qualified to service that system.

7. Isolate the Facility AGSS Terminal

Test the wall or ceiling AGSS terminal according to facility policy.

Possible isolation methods include:

Expected:

Do not return the machine to the original location merely because it operates normally at another outlet. The defective terminal must also be documented and corrected.

8. Inspect the External AGS Assembly

Visually inspect the receiving-system assembly for:

Do not perform internal disassembly or attempt to clean an inaccessible restriction unless permitted by approved service procedures.

Expected: The AGS assembly is complete, secure, clean, and undamaged.

A suspected blocked or missing internal filter requires replacement of the appropriate AGS component or evaluation by qualified service personnel.

9. Perform the Approved System Checkout

After correcting the suspected cause:

Expected: The machine completes its checkout with stable scavenging flow and no evidence of waste-gas leakage.

If all testing passes, return the device to service according to department policy.

If the Problem Persists

If the hose, connections, routing, accessories, and facility AGSS terminal have been verified, the fault may involve the integrated receiving system, filter, mounting assembly, or another internal scavenging component.

The device should be:

A machine must not be returned to clinical use with an unstable flow indicator, persistent anesthetic odor, failed system test, or unresolved evacuation concern.

Knowing when to stop external troubleshooting and escalate the equipment is proper Clinical Engineering practice.

Clinical Use Tip

Never troubleshoot an unresolved scavenging problem during active anesthesia unless directed by the anesthesia provider as part of an immediate patient-safety response. Move the patient to another verified anesthesia system before extended testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported weak scavenging, and the Perseus AGS flow indicator remained below the restricted range."

Cause

What was observed during troubleshooting.

Example:
"The AGSS suction hose was compressed behind the rear caster, restricting waste-anesthetic-gas evacuation."

Resolution

What action was taken.

Example:
"Rerouted and secured the hose, verified the flow indicator remained in the normal range, completed the system checkout, and returned the machine to service."

Helpful Details to Include (If Known)

Final Thought

Scavenging problems can expose personnel to waste anesthetic gas and may originate from either the machine or the facility AGSS. Logical isolation, safe patient transfer, appropriate escalation, and complete CCR documentation protect both patients and staff.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide