On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
A battery-backed Polaris 600 installation does not charge properly or provides less backup operating time than expected.
Step-by-Step Troubleshooting
1. Protect the Patient and Confirm Backup Lighting
Do not depend on questionable battery-backed surgical lighting during a procedure. Confirm adequate alternate lighting and facility emergency-power coverage before testing the affected backup function.
Expected outcome: Patient care is protected independently of the suspected battery problem.
2. Confirm the Installation Actually Uses Battery Backup
Identify whether the reported battery is part of the Polaris 600 installation, an external emergency-power accessory, or another facility-supported component. Do not assume every Polaris 600 configuration contains an internal user-serviceable battery.
Expected outcome: The battery complaint is associated with the correct component and system configuration.
3. Confirm the Exact Symptom
Determine whether the battery does not charge, indicates an abnormal state, fails to support the light during a permitted test, or provides reduced runtime. Record any available indicators.
Expected outcome: The complaint is reproduced or clearly defined before parts are replaced.
4. Verify Normal AC Supply
Confirm the lighting system and any associated charging equipment have their required normal AC supply. Check accessible power connections and verify that the room is not experiencing a facility electrical problem.
Expected outcome: Normal AC charging power is present. If restoring external power restores charging behavior, verify operation and troubleshooting can stop.
5. Inspect Accessible Battery or Backup-Power Connections
Where the installation provides accessible battery connections or an external backup-power module, inspect for loose connections, damage, corrosion, contamination, or abnormal heat. Do not open sealed battery assemblies or fixed electrical enclosures outside authorized service procedures.
Expected outcome: Accessible connections are secure and physically acceptable.
6. Allow Normal Charging Opportunity
If the battery was recently discharged or the system was disconnected from AC power, allow it to charge under normal approved operating conditions before judging runtime.
Do not invent or apply an unsupported charging interval.
Expected outcome: Battery status improves after normal charging. If charging and backup operation become normal, troubleshooting can stop after verification.
7. Evaluate Backup Operation Safely
When permitted by facility procedure and manufacturer documentation, test backup operation without placing a patient or active procedure at risk. Observe whether lighting remains available and whether abnormal indications occur.
Expected outcome: Backup operation is stable for the approved functional test. If performance is satisfactory, troubleshooting is complete.
8. Escalate Persistent Battery Problems
If charging power and accessible connections are normal but the battery remains unable to charge or support its intended backup function, remove the backup function from clinical reliance.
Expected outcome: The battery or backup-power system is referred for qualified evaluation rather than being used unreliably.
If the Problem Persists
External charging and connection problems have been ruled out. Remaining causes may include battery deterioration, charging circuitry, backup-power electronics, configuration, or facility emergency-power infrastructure.
The affected equipment should be:
- Removed from service when required backup lighting cannot be assured
- Labeled Out of Service
- Sent for repair or appropriate bench/on-site evaluation
- Evaluated using applicable manufacturer documentation and approved test equipment
- Repaired, battery-serviced, or configured only by qualified personnel
Complete functional and electrical safety testing appropriate to the repair before return to service. Stopping after external causes are eliminated prevents unsafe assumptions.
Clinical Use Tip
Do not perform backup-power testing during a case unless adequate independent surgical illumination is already available.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported that the Polaris 600 backup lighting did not remain available during a facility power interruption test."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found that the associated backup-power connection was not fully seated."
Resolution
What action was taken.
Example:
"Secured the accessible connection and verified normal charging indication and successful backup-lighting functional operation before return to service."
Helpful Details to Include (If Known)
- Type of backup-power configuration
- Charge indication
- Normal AC power status
- Whether the battery was recently discharged
- Accessible battery connections inspected
- Backup operation observed
- Abnormal heat, swelling, odor, or damage
- Results before and after correction
- Final device status
Final Thought
Treat backup lighting as a patient-safety function. Verify configuration, charging power, connections, and controlled backup operation before assuming battery failure, and escalate any unreliable backup system with clear documentation.
That is successful troubleshooting.