Drager Polaris 600

Unexpected Shutdown, Freeze, or Restart

On this page

Asset Type

Surgical Light

Manufacturer

Drager

Model

Polaris 600

What This Guide Helps With

The Polaris 600 unexpectedly turns off, stops responding, or restarts during operation.

Step-by-Step Troubleshooting

1. Immediately Maintain Surgical Illumination

If the failure occurs during a procedure, establish adequate alternate surgical lighting immediately. Do not continue depending on a light that has shut down, frozen, or restarted unexpectedly.

Expected outcome: The surgical field remains safely illuminated independently of the affected unit.

2. Document the Exact Failure

Determine whether illumination stopped, controls froze while light output remained, the system restarted automatically, or the entire installation lost power. Record what was happening immediately before the event.

Expected outcome: The failure mode is clearly described and can be compared with power, movement, or environmental conditions.

3. Check Facility Power

Verify stable facility power to the installation and determine whether other equipment in the room experienced a simultaneous power interruption.

Expected outcome: Facility power is confirmed stable. If the event coincided with an external power interruption and operation remains normal after restoration, document and verify before stopping troubleshooting.

4. Inspect Accessible Connections

Check accessible power and communication connections for looseness, damage, or strain. Observe whether movement of the light or suspension is associated with loss of operation, but do not repeatedly provoke a known intermittent shutdown.

Expected outcome: No external connection issue is found. If a secure external connection corrects the problem, verify stable operation and stop troubleshooting.

5. Inspect for Environmental or Physical Concerns

Look for fluid intrusion, excessive contamination, blocked ventilation openings where applicable, unusual heat, odor, visible damage, or evidence of impact.

Expected outcome: No environmental or physical condition is present that would make continued operation unsafe.

6. Perform a Controlled Restart

With the device removed from clinical dependence, complete a normal power restart and observe the entire startup process.

Expected outcome: The system starts normally without freezing or restarting.

7. Exercise Normal Controls and Movement

Operate the light through normal external controls and positioning while watching for interruption, flicker, control freeze, or restart.

Do not intentionally stress cables or suspension components.

Expected outcome: Operation remains stable throughout normal use. If the issue cannot be reproduced and all verification is satisfactory, follow local policy regarding intermittent failures and documentation.

8. Remove From Service for Repeated or Unexplained Failure

A repeated or unexplained shutdown, freeze, or restart can compromise surgical illumination and should not be dismissed simply because the light later operates normally.

Expected outcome: Unreliable equipment is removed from clinical use and escalated appropriately.

If the Problem Persists

External power, connections, and environmental causes have been evaluated. Remaining categories may include internal power electronics, control hardware, communication, software, thermal conditions, or installation wiring.

The affected equipment should be:

Following repair, verify startup, stable illumination, control response, movement, and any applicable electrical safety requirements before return to service.

Clinical Use Tip

Treat an unexplained intermittent loss of surgical illumination as clinically significant even when the light operates normally afterward.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"OR staff reported that one Polaris 600 light head shut off unexpectedly when it was repositioned."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an accessible external connection associated with the affected assembly was loose and movement caused intermittent operation."

Resolution

What action was taken.

Example:
"Secured the connection and verified repeated positioning, stable illumination, and normal control response before return to service."

Helpful Details to Include (If Known)

Final Thought

Intermittent shutdowns require caution because they may directly affect the surgical field. Verify external power, connections, movement-related behavior, and environment first, then remove unreliable equipment from service and document findings clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide