Drager Primus

Bellows Not Cycling Correctly

On this page

Asset Type

Ventilator

Manufacturer

Drager

Model

Primus

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting Drager Primus ventilators when the bellows fails to cycle correctly. A bellows that does not rise and fall properly can indicate airflow obstruction, gas delivery issues, or mechanical problems, potentially compromising patient ventilation. Most causes are external or easily verifiable before considering internal repair.

Step-by-Step Troubleshooting

Verify Power and Initialization

Check Gas Supply Lines

Inspect the Bellows Assembly and Circuit

Review Ventilator Settings

Check Pressure Relief Valves and Filters

Perform a Pre-Use Self-Test

If the Problem Persists

External checks have been completed and no issues are found. Bellows still fails to cycle correctly.

Action:

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“User reports the bellows does not rise or fall properly during ventilation.”

Cause

What was observed during troubleshooting.

Example:
“Observed that gas supply lines were connected but the exhalation valve had minor obstruction preventing normal bellows cycling.”

Resolution

What action was taken.

Example:
“Cleared obstruction from exhalation valve and verified bellows cycled correctly during pre-use self-test. If unresolved, ventilator removed from service and sent for vendor evaluation.”

Helpful Details to Include

Final Thought

Always prioritize patient safety over troubleshooting. Start with logical, external checks before assuming internal failure. Accurate documentation supports efficient repair and prevents repeated downtime.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide