Drager Primus IE

Ventilator Bellows Not Cycling

On this page

Asset Type

Anesthesia Machine

Manufacturer

Drager

Model

Primus IE

What This Guide Helps With

Identifies causes for bellows failing to rise/fall, including power, circuit, or valve issues, before patient exposure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety

2. Verify Power Supply

3. Check Oxygen/Compressed Gas Supply

4. Inspect Breathing Circuit Connections

5. Examine Ventilator Settings

6. Check for Active Alarms or Obstructions

7. Perform a System Reset

If the Problem Persists

All external checks have been completed; internal pneumatic or mechanical failure is likely. Remove from service, label as Out of Service, and send for repair or bench evaluation. Document that further troubleshooting requires qualified repair personnel.

Clinical Use Tip

Never attempt to ventilate a patient with a non-cycling bellows. Always switch to a functioning anesthesia machine or manual ventilation before troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Bellows not cycling during pre-use check; patient unable to ventilate."

Cause

What was observed during troubleshooting.

Example:
"Breathing circuit properly connected; gas supply and power verified; issue persisted. Likely internal ventilator fault."

Resolution

What action was taken.

Example:
"Device removed from service, labeled Out of Service, sent for repair."

Helpful Details to Include (If Known)

Final Thought

Patient safety is the priority. Bellows not cycling is a critical ventilator failure; stepwise troubleshooting ensures simple causes are addressed, but internal repair should be escalated promptly. Accurate documentation aids future preventive actions.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide