Drager Resuscitaire Series

Radiant Heater Failure

On this page

Asset Type

Infant Warmer

Manufacturer

Drager

Model

Resuscitaire Series

What This Guide Helps With

Troubleshooting absent or inadequate radiant heat output caused by power, mode, alarm, control, obstruction, or heater assembly faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Remove the infant from the affected warmer and transfer care to an approved alternate warming device according to clinical protocol.

Do not test a suspected heater failure while an infant depends on the device for thermal support.

Expected outcome: The infant is safely supported on a working warmer before troubleshooting begins.

Why it matters: Failure of radiant heat can quickly compromise neonatal temperature management.

2. Confirm the Reported Heater Failure

With the device out of clinical use, verify the reported condition:

The Resuscitaire warmer is designed with adjustable heater output from 0% to 100% in 10% increments and includes controller heating alarms such as System Fail and Power Fail.

Expected outcome: The complaint is reproduced or the reported alarm/behavior is documented.

If the heater operates normally and no alarm returns: Complete appropriate functional verification before returning the device to service, then stop.

3. Inspect for Visible Damage or Unsafe Conditions

Inspect the heater head, reflector area, control panel, column, power cord, and plug for:

Do not touch the radiant heater surface immediately after operation; allow adequate cooling time.

Expected outcome: No visible evidence of overheating, damage, or contamination is present.

If damage, smoke, burning odor, or abnormal heat is found: Stop troubleshooting, remove the unit from service, and send it for repair evaluation.

4. Verify AC Power and Power-Cord Connection

Confirm the power cord is firmly seated at the device and connected to a known-working hospital-grade outlet.

Check whether:

The Resuscitaire warmer requires AC power for heater operation; product specifications list 120 V or 220–240 V configurations depending on the model.

Expected outcome: The warmer receives stable AC power without a power-fail alarm.

If power restoration resolves the issue: Complete functional verification and stop.

5. Verify Heater Mode and Selected Output

Confirm the warmer is placed into the appropriate testable operating mode according to facility procedure and that heater output is intentionally set above 0%.

Observe whether:

Expected outcome: The control panel acknowledges a heat demand and displays an appropriate selected heater output.

If the heater was unintentionally set to 0% or an incorrect mode and normal heating returns after correction: Complete functional verification and stop.

6. Check for Alarm Conditions That May Interrupt Heating

Observe the alarm indicators and display while heat is commanded. Document any active or repeated indication, including:

Do not repeatedly silence or reset alarms simply to continue operation without identifying the cause.

Expected outcome: No unresolved alarm prevents normal heater operation.

If a persistent alarm occurs with loss of heating: Remove the device from service and proceed to repair evaluation.

7. Inspect the Heater Path and Physical Position

Verify that the heater head is properly positioned over the mattress area and that nothing blocks radiant heat delivery, including:

Do not disassemble or clean internal heater components during basic troubleshooting.

Expected outcome: The radiant heat path is unobstructed and the heater is correctly positioned.

If removing an obstruction restores normal warming: Complete functional verification and stop.

8. Assess Whether Radiant Heat Is Produced

With the unit empty and safely positioned for testing, command heater output and allow an appropriate observation period according to facility policy.

Observe from a safe distance whether:

The Resuscitaire uses a radiant heater and reflector arrangement intended to distribute heat across the mattress area.

Expected outcome: The warmer produces consistent radiant heat without abnormal alarms, odor, or visible damage.

If indicated output is present but radiant heat is absent or inadequate: Suspect a heater assembly, control, wiring, or internal power-delivery fault and stop further external troubleshooting.

9. Perform Final Functional Verification Only If the Problem Was Corrected

When an external issue has been identified and corrected, perform the required operational and safety verification for the warmer according to departmental procedure before returning it to use.

Verify that:

Expected outcome: The infant warmer is confirmed safe and functional for clinical use.

If verification fails: Remove the warmer from service and send it for repair evaluation.

If the Problem Persists

If stable AC power, proper heater mode, selected output, unobstructed heater path, visible condition, and alarm behavior have been checked but the warmer still does not provide reliable radiant heat, common external causes have been ruled out.

The failure is likely related to an internal heater element, heater control circuitry, internal connection, sensor/control fault, or power-delivery component.

The device should be:

Knowing when to stop external troubleshooting and escalate a neonatal warming failure is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a radiant heater failure while an infant is relying on the warmer. Move the patient to a known-working warming device first, then evaluate the affected unit only when it is safe and unoccupied.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Drager Resuscitaire powered on but did not provide radiant heat during setup for newborn care."

Cause

What was observed during troubleshooting.

Example:
"Verified stable AC power and selectable heater output, but the heater produced no detectable warming and the fault persisted with the unit empty."

Resolution

What action was taken.

Example:
"Removed the infant warmer from service, applied an Out of Service label, and routed the unit for bench evaluation of the radiant heater and internal heater-control system."

Helpful Details to Include (If Known)

Final Thought

Radiant heater failure on an infant warmer is a patient-safety concern that requires a methodical response: protect the infant first, rule out external causes, verify the failure safely, and escalate promptly when dependable heat cannot be confirmed. Accurate CCR documentation preserves the troubleshooting trail and supports appropriate repair decisions.

That is successful troubleshooting.