Drager Resuscitaire Series

Suction System Failure

On this page

Asset Type

Infant Warmer

Manufacturer

Drager

Model

Resuscitaire Series

What This Guide Helps With

Troubleshooting absent, weak, or inconsistent integrated suction caused by tubing, collection container, regulator, or supply-source issues.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

If the warmer is being used during newborn care and suction is unavailable or unreliable, notify clinical staff immediately and provide a functioning alternate suction source.

Do not continue troubleshooting the suction system while it may be needed for active patient care.

Expected outcome: The patient has reliable suction available before equipment troubleshooting begins.

2. Confirm the Reported Suction Failure

Confirm that the unit is equipped with the integrated resuscitation/suction module and identify the reported symptom:

The Resuscitaire suction circuit is specified as adjustable from 0 to 150 mmHg and uses a 40 ml disposable collection container on equipped configurations.

Expected outcome: The exact suction-related failure is identified before parts are exchanged or the device is removed from service.

3. Inspect the Suction Patient Connection and Tubing

With the device out of patient use, inspect the suction tubing and patient connection for:

Replace questionable disposable tubing or suction accessories with compatible components, then retest.

Expected outcome: Tubing is clear, secure, and properly connected. If normal suction returns, document the replaced disposable accessory and stop troubleshooting.

4. Inspect the Collection Container and Lid/Seal

Check the disposable collection container for:

Replace the disposable collection container when damaged, contaminated, full, or suspected of leaking.

Expected outcome: The collection container is empty or replaced, correctly installed, and able to maintain a vacuum seal. If suction returns, document the cause and stop.

5. Verify Suction Adjustment and Gauge Response

Using an appropriate test setup and not a patient, turn the suction control through its adjustment range while observing the vacuum gauge.

Briefly occlude the test tubing tip to determine whether vacuum develops and whether the gauge responds smoothly.

Expected outcome: Vacuum increases with adjustment and the gauge indicates suction. If the control does not change vacuum or the gauge remains at zero, continue troubleshooting.

6. Check the Resuscitation Gas Supply Source

Confirm that the connected gas source required for the resuscitation module is present, properly connected, and available at the warmer.

Inspect external supply connections for:

Do not attempt internal pneumatic adjustment.

Expected outcome: The external gas source is connected and available. If restoring the gas supply restores suction, document the external supply issue and stop.

7. Isolate the Failure to the Suction Path

If available and approved for testing, install known-good compatible suction tubing and a new collection container, then repeat the occlusion test.

If suction returns, the removed disposable component was the likely cause.

If suction remains absent or weak despite known-good external components and an available gas supply, the failure is likely within the suction regulator, gauge, internal pneumatic path, or resuscitation module.

Expected outcome: External disposable and supply-related causes are either corrected or ruled out.

8. Inspect for Contamination, Leakage, or Physical Damage

Inspect the accessible suction module area for:

Do not continue functional testing if contamination may have entered the device’s internal pneumatic components.

Expected outcome: Any visible damage or contamination is identified and documented for repair evaluation.

If the Problem Persists

If suction remains absent, weak, nonadjustable, or unreliable after verifying tubing, disposable collection container, external connections, and available gas supply, common external causes have been ruled out.

The issue is likely internal to the suction regulator, gauge, pneumatic pathway, or resuscitation module.

The device should be:

Do not attempt internal pneumatic repair, regulator adjustment, or component-level service during floor troubleshooting. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a suction failure while the device is supporting active newborn care. Ensure a working alternate suction source is available before removing the Resuscitaire from service or performing testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Labor and Delivery reported that the Drager Resuscitaire integrated suction produced little to no vacuum during setup for newborn care."

Cause

What was observed during troubleshooting.

Example:
"Suction remained weak after replacement of the disposable collection container and tubing, with external gas supply present and no adequate vacuum response during occlusion testing."

Resolution

What action was taken.

Example:
"Removed the infant warmer from service, labeled it Out of Service, and sent it for bench evaluation of the integrated suction/resuscitation module."

Helpful Details to Include (If Known)

Final Thought

An infant warmer suction failure can directly affect readiness during newborn care. Start with patient safety, then rule out tubing, container, connection, and gas-supply issues before escalating suspected internal pneumatic failure. Clear CCR documentation supports safe repair follow-up and prevents unnecessary repeat troubleshooting.

That is successful troubleshooting.