On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots failed or intermittent user-interface controls and indicators caused by contamination, damage, power, accessories, software state, or service-level faults.
Step-by-Step Troubleshooting
1. Protect the Patient From an Unreliable User Interface
Do not continue patient use if essential settings cannot be viewed or changed, alarms cannot be reliably recognized, or the control interface is intermittent.
Transfer the patient to another verified ventilator before troubleshooting.
Expected outcome: The patient is supported by reliable equipment and the Savina 300 can be safely evaluated off-patient.
2. Confirm the Exact Interface Failure
Determine which function is affected:
- Display blank or distorted
- Display freezes
- Rotary knob does not turn or select reliably
- Touch control does not respond
- Alarm light does not illuminate
- Controls respond intermittently
- Multiple interface elements fail together
Document whether ventilation continues during the interface failure.
Expected outcome: The affected control or indicator is clearly identified.
3. Verify Basic Power and Startup
Confirm stable AC power and restart the ventilator under controlled off-patient conditions.
Observe whether all startup indications, display elements, and controls initialize normally.
Expected outcome: The interface initializes and remains stable. If a power interruption caused the problem and stable power restores operation, continue to final verification.
4. Inspect for Physical Damage
Inspect the display, rotary knob, control panel, alarm indicator, housing, and surrounding surfaces for:
- Cracks
- Impact damage
- Loose controls
- Missing pieces
- Fluid intrusion
- Cleaning-related deterioration
Do not return equipment with damaged essential controls to service.
Expected outcome: The user interface is mechanically intact. Visible damage affecting safe operation requires removal from service.
5. Check for Contamination or Moisture
Inspect control surfaces for residue, dried cleaning solution, fluid, adhesive material, or contamination that may interfere with control operation.
Clean only according to approved procedures and allow surfaces to dry as required.
Expected outcome: Controls are clean and dry. If external contamination caused the problem, verify repeated reliable operation before proceeding.
6. Remove Nonessential External Connections
Disconnect nonessential communication cables and peripherals, then restart and reevaluate the interface.
Reconnect items individually if the problem disappears.
Expected outcome: The controls remain functional with essential connections only. If a peripheral or cable correlates with the fault, evaluate that item separately.
7. Exercise Each Control Off-Patient
Operate the rotary knob and accessible control inputs repeatedly through normal user functions without entering unauthorized menus.
Check for missed inputs, unintended double activations, sticking, or intermittent response.
Expected outcome: Every essential control responds consistently. If any control remains intermittent, the ventilator should not return to service.
8. Verify Display and Alarm Indicators
During an approved functional check, verify the display remains readable and the visual and audible alarm indications operate as expected.
Do not create unsafe test conditions merely to trigger an alarm; use approved testing methods.
Expected outcome: Display and alarm indications are clearly visible and operate reliably.
9. Perform Extended Functional Verification
Operate the ventilator on a test lung and observe the interface during changes to permitted test settings, alarm testing, and normal operation.
Watch for freezing, flickering, delayed control response, or spontaneous restart.
Expected outcome: The interface remains continuously stable and responsive. Troubleshooting can stop when all essential controls and indicators pass verification.
10. Escalate Persistent Interface Failure
If an essential display, control, or alarm indicator remains failed or intermittent after external causes have been ruled out, remove the ventilator from service.
Expected outcome: The ventilator is either verified fully controllable and observable or sent for service-level evaluation.
If the Problem Persists
External power, contamination, physical condition, accessories, and simple restart causes have been ruled out. Remaining possibilities may involve the display assembly, control interface, internal communication, control electronics, software, alarm-indicator circuitry, or another service-level fault.
The ventilator should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate Drager documentation and approved test equipment
- Repaired or configured only by qualified personnel
After repair, verify every essential control, display function, alarm indicator, ventilation function, and applicable electrical-safety requirement before return to service.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
A ventilator that continues delivering breaths is still unsafe if staff cannot reliably view settings, change controls, or recognize alarms.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Respiratory Therapy reported that the Savina 300 rotary control responded intermittently during setup."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found dried cleaning residue around the external rotary control that interfered with consistent operation."
Resolution
What action was taken.
Example:
"The control area was cleaned using the approved external cleaning procedure and repeated control, alarm, display, and test-lung functional checks were completed successfully."
Helpful Details to Include (If Known)
- Exact control or indicator affected
- Whether ventilation continued
- AC or battery operation
- Physical damage observed
- Moisture or cleaning residue
- External accessories connected
- Restart result
- Alarm-light and speaker behavior
- Results of repeated control testing
- Final device status
Final Thought
The user interface is part of the ventilator's safety system. Protect the patient, verify simple external causes before assuming electronics failure, require reliable operation of every essential control and alarm indicator, escalate unresolved faults, and document final verification clearly.
That is successful troubleshooting.