Drager Vamos 2

Measurement Drift or Calibration Issue

On this page

Asset Type

Anesthesia Monitor

Manufacturer

Drager

Model

Vamos 2

What This Guide Helps With

Troubleshooting drifting gas measurements, unstable readings, or calibration-related issues before sending the monitor for repair.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot a suspected measurement accuracy issue on an active patient unless safe alternate monitoring is already in place.

Action:

Expected outcome: Patient safety is protected before troubleshooting begins.

Why it matters: A drifting anesthesia gas or CO₂ measurement can affect ventilation assessment, agent delivery awareness, and clinical decision-making.

If this resolves the immediate safety concern, continue troubleshooting off-patient.

2. Verify the Reported Measurement Problem

Confirm exactly what is drifting or failing calibration.

Action:

Expected outcome: The reported issue is clearly identified before parts or accessories are changed.

Why it matters: “Calibration issue” can be caused by the monitor, sampling line, water trap, gas source, contamination, or incorrect setup.

3. Check for Active Alarms or Status Messages

Look for occlusion, sampling, calibration, zeroing, or gas measurement messages.

Action:

Expected outcome: Any displayed alarm helps narrow the cause.

If the monitor shows a sampling system or occlusion alarm, troubleshoot the sampling path before assuming calibration failure.

4. Inspect the Sampling Line

Check the sample line from the patient connection to the Vamos 2.

Action:

Expected outcome: Gas readings stabilize after replacing a restricted or contaminated sample line.

Why it matters: Partial restriction or contamination can cause delayed response, incorrect values, or drifting measurements.

If readings stabilize, document the faulty sample line and return the monitor to service after verification.

5. Check the Water Trap

Inspect the water trap for fluid, contamination, improper seating, or age-related issues.

Action:

Expected outcome: The monitor samples normally and measurements become stable.

Why it matters: A leaking or saturated water trap can affect sample flow and measurement accuracy.

If the issue clears after replacing or reseating the water trap, stop and document the accessory-related cause.

6. Allow Proper Warm-Up Time

Confirm the monitor has had enough time to stabilize after power-up.

Action:

Expected outcome: Readings become stable after normal warm-up.

Why it matters: Gas analyzers may require stabilization time before accurate measurement is available.

If the drift disappears after warm-up, document that no fault was duplicated after stabilization.

7. Verify the Sample Gas Source

If bench testing, use a known-good gas source or simulator appropriate for the parameter being checked.

Action:

Expected outcome: The Vamos 2 reading matches the expected gas value within acceptable facility or manufacturer tolerance.

Why it matters: A bad test source can make a good monitor appear out of calibration.

8. Compare Against Another Known-Good Monitor

Use a second verified analyzer when possible.

Action:

Expected outcome: The Vamos 2 either agrees with the reference monitor or shows a repeatable offset/drift.

Why it matters: Comparison testing helps separate a true device issue from a setup or gas source issue.

If the Vamos 2 agrees with the reference monitor, document no fault found and include the test conditions.

9. Check for Leaks in the Sampling Path

A small leak can dilute the sampled gas and cause low, unstable, or drifting values.

Action:

Expected outcome: Readings stabilize when the sampling path is sealed.

Why it matters: Leaks can mimic calibration drift by allowing room air into the gas sample.

10. Check for Contamination or Residual Moisture

Moisture, residue, or contaminated accessories can cause slow response and measurement instability.

Action:

Expected outcome: The analyzer clears and readings stabilize.

Why it matters: Contamination in the sampling path can cause false drift even when the internal analyzer is functioning.

11. Perform User-Level Zeroing or Calibration Check Only if Allowed

Use only facility-approved and manufacturer-supported procedures available to Clinical Engineering.

Action:

Expected outcome: The device passes verification or clearly fails in a repeatable way.

Why it matters: Clinical Engineering should verify performance without creating an undocumented or unauthorized calibration change.

12. Power Cycle and Recheck

If the issue is intermittent, power cycle the monitor and retest.

Action:

Expected outcome: The issue either clears or returns consistently.

Why it matters: A repeatable failure is stronger evidence of an internal measurement or calibration problem.

13. Determine Whether the Device Can Be Returned to Service

Only return the monitor to service if the readings are stable and verified.

Action:

Expected outcome: The monitor is safe to return to clinical use.

If measurement drift remains, do not return the device to service.

If the Problem Persists

If the Vamos 2 continues to show measurement drift, unstable gas values, failed zeroing, or failed calibration verification after external causes are ruled out, the issue is likely internal.

Remove the device from service.

Label it Out of Service.

Send it for repair, bench evaluation, or vendor/service-level calibration.

Common external causes such as sample line restriction, water trap issues, leaks, contamination, warm-up time, and bad test gas should be ruled out first. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Do not troubleshoot questionable gas measurements during an active case unless patient safety is protected first. Move monitoring to a known-good device before testing, replacing accessories, or performing verification checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia reported the Drager Vamos 2 gas readings were drifting and did not match expected CO₂ and agent values during use."

Cause

What was observed during troubleshooting.

Example:
"Inspection found a contaminated water trap and moisture in the sampling line causing unstable gas measurement readings."

Resolution

What action was taken.

Example:
"Replaced the water trap and sample line, allowed the monitor to warm up, verified stable readings with a known-good test setup, and returned the device to service."

Helpful Details to Include (If Known)

Final Thought

Measurement drift should be treated as a patient safety concern until proven otherwise. Start with the sampling path, water trap, leaks, accessories, warm-up, and test setup before suspecting internal analyzer failure. If readings cannot be verified, remove the monitor from service and document the complaint, cause, and resolution clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide