Edwards Lifesciences HemoSphere Advanced Monitoring System

Cardiac Output Or Advanced Parameter Values Missing / Unreliable

On this page

Asset Type

Hemodynamic Monitor

Manufacturer

Edwards Lifesciences

Model

HemoSphere Advanced Monitoring System

What This Guide Helps With

Troubleshooting missing, unstable, or questionable cardiac output and advanced hemodynamic values caused by setup, sensor, cable, module, or signal issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Do not troubleshoot the HemoSphere system in a way that delays patient care or compromises active monitoring.

Action:

Expected outcome: Patient care continues safely while Clinical Engineering evaluates the equipment side of the issue.

Why it matters: Advanced hemodynamic values may influence treatment decisions. Unreliable data should not be treated as accurate until verified.

Verify the Reported Problem

Confirm exactly which value is missing or unreliable.

Action:

Ask staff which parameter is affected, such as:

Check whether the issue is:

Expected outcome: The failure is narrowed to a specific parameter, module, cable, sensor, or setup path.

Why it matters: “Cardiac output not working” can be caused by several different external issues depending on which monitoring technology is being used.

Confirm the Correct Monitoring Setup Is Selected

Verify that the HemoSphere monitor is set up for the intended monitoring mode and connected technology.

Action:

Expected outcome: The monitor is in the correct mode and capable of displaying the requested parameter.

Why it matters: Some advanced values only appear when the correct Edwards module, pressure cable, sensor, or oximetry setup is connected and configured.

Check the Patient Cable, Module, and Connector Seating

Inspect all external connections before assuming a monitor failure.

Action:

Inspect connectors for:

Expected outcome: The module and cable are fully seated and recognized by the monitor.

Why it matters: A loose or damaged cable can cause missing values, intermittent readings, or unreliable hemodynamic calculations.

Check for Sensor or Disposable Setup Issues

Confirm that the correct disposable sensor or patient connection is being used and properly connected.

Action:

Expected outcome: The disposable sensor path is intact and suitable for measurement.

Why it matters: Many missing or unreliable values are caused by sensor setup, line condition, or disposable issues rather than the monitor itself.

Verify Pressure Waveform Quality If Pressure-Derived Values Are Affected

If cardiac output or advanced parameters are calculated from arterial pressure, confirm the waveform is usable.

Action:

Look for:

Expected outcome: The pressure waveform is clean, stable, and clinically appropriate.

Why it matters: Pressure-derived cardiac output and advanced parameters depend on waveform quality. A poor waveform can produce missing or unreliable values.

If waveform correction resolves the issue, document the finding and stop.

Confirm Required Patient Data Is Entered Correctly

Some calculated values require accurate patient demographics or setup data.

Action:

Check that required patient information is entered correctly, such as:

Expected outcome: The monitor has the required information to calculate indexed or advanced values.

Why it matters: Missing or incorrect patient data can prevent certain calculated values from appearing or can make displayed values unreliable.

Check Alarm, Message, and Status Notifications

Review any on-screen messages that explain why a value is unavailable.

Action:

Expected outcome: The displayed message identifies the likely affected accessory, module, sensor, or setup condition.

Why it matters: The monitor may already be identifying the failed portion of the measurement chain.

Swap External Accessories With Known-Good Parts

Use known-good external parts when available.

Action:

If clinically safe and approved, test with:

Expected outcome: The issue follows a cable, module, or sensor, or remains with the monitor.

Why it matters: Controlled accessory swaps help separate monitor failure from external connection or disposable failure.

If a known-bad accessory is identified, remove it from service and stop.

Restart the Monitor Only When Clinically Safe

If the display is not updating, values appear frozen, or the system behavior is abnormal, perform a controlled restart only when patient monitoring can safely continue another way.

Action:

Expected outcome: Temporary software or communication issues clear after restart.

Why it matters: A restart can correct a temporary system lockup, but it should never interrupt critical patient monitoring without a safe backup.

Check Power and Battery Status

Confirm the monitor has stable power.

Action:

Expected outcome: The monitor has stable AC power and is not operating under a power-related fault condition.

Why it matters: Unstable power can contribute to system errors, unexpected shutdowns, or unreliable module communication.

Check Network or Data Export Only If the Values Are Present Locally

If the values appear correctly on the HemoSphere monitor but are missing in the EMR, central display, or data export system, separate the issue from bedside measurement.

Action:

If present locally but missing elsewhere, check:

Expected outcome: The problem is identified as a connectivity/export issue rather than a measurement issue.

Why it matters: A value can be clinically valid on the monitor but still fail to export because of network or interface problems.

Perform a Basic Functional Check Off Patient

If the issue cannot be resolved at the bedside, remove the monitor from patient use and test it in a safe non-patient environment.

Action:

Expected outcome: Clinical Engineering determines whether the issue is accessory-related, configuration-related, or likely internal to the monitor.

Why it matters: Bench evaluation prevents unnecessary patient disruption and supports better repair decisions.

If the Problem Persists

If the cardiac output or advanced parameter values remain missing, unstable, frozen, or clinically unreliable after checking setup, cables, modules, sensors, waveform quality, patient data, power, and restart behavior, the likely cause is internal monitor failure, module failure, software fault, or a problem requiring manufacturer-level service.

The device should be:

Knowing when to stop is proper troubleshooting. Do not keep returning a monitor to patient use when advanced hemodynamic values cannot be trusted.

Clinical Use Tip

Do not troubleshoot missing or unreliable hemodynamic values on an active patient unless it is safe and clinically approved. If values are being used for treatment decisions, have staff move to another monitoring method or another HemoSphere monitor before equipment troubleshooting continues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Edwards HemoSphere monitor was not displaying cardiac output and advanced hemodynamic values during patient monitoring."

Cause

What was observed during troubleshooting.

Example:
"Found arterial waveform quality was poor due to an external pressure setup issue, preventing reliable pressure-derived parameter calculation."

Resolution

What action was taken.

Example:
"Clinical staff corrected the pressure setup, waveform returned to normal, cardiac output and advanced parameter values repopulated, and monitor remained in service."

Helpful Details to Include (If Known)

Final Thought

Troubleshooting missing or unreliable HemoSphere cardiac output values should follow the signal path from patient safety, setup, waveform quality, cables, modules, sensors, and power before assuming internal failure. If the data cannot be trusted after external causes are ruled out, remove the monitor from service and document the complaint, cause, and resolution clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide