Edwards Lifesciences HemoSphere Advanced Monitoring System

Invasive Pressure Waveform Absent, Dampened, or Inaccurate

On this page

Asset Type

Hemodynamic Monitor

Manufacturer

Edwards Lifesciences

Model

HemoSphere Advanced Monitoring System

What This Guide Helps With

Troubleshooting missing, dampened, inaccurate, or unreliable invasive pressure waveforms caused by setup, cable, sensor, leveling, zeroing, or connection issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Do not troubleshoot the pressure monitoring setup on an unstable patient unless clinical staff confirms it is safe.

Action:

Expected outcome: Patient care continues safely while the equipment issue is isolated.

Why it matters: An inaccurate arterial, CVP, or PA pressure waveform can affect clinical decisions.

Verify the Reported Symptom

Confirm whether the issue is a total loss of waveform, dampened waveform, incorrect numeric value, or intermittent dropout.

Action:

Expected outcome: The failure mode is clearly identified before parts are swapped.

Check the Pressure Cable Connection

The HemoSphere system uses pressure cable and sensor/transducer connections for invasive pressure monitoring, so loose or incorrect cable connections can cause missing or unreliable values. Edwards setup material references connecting the HemoSphere pressure cable, sensor/transducer, leveling, and zeroing as part of pressure monitoring setup.

Action:

Expected outcome: The pressure channel detects the cable and sensor properly.

If the waveform returns after reseating or replacing the cable, document the connection or cable issue and stop.

Check the Disposable Pressure Sensor or Transducer

A failed, expired, damaged, wet, or incorrectly connected disposable sensor can cause no waveform or inaccurate readings.

Action:

Expected outcome: A known-good sensor provides a stable pressure signal.

Verify Tubing, Stopcocks, and Line Setup

Dampened or inaccurate invasive pressure waveforms are commonly caused by line setup issues rather than monitor failure.

Action:

Expected outcome: The waveform becomes sharper, more responsive, and clinically believable.

If the waveform improves after correcting tubing, stopcock, or bubble issues, document the external cause and stop.

Verify Pressure Bag and Flush System

For fluid-filled invasive pressure monitoring, an underinflated pressure bag or restricted flush setup can cause dampened waveforms.

Action:

Expected outcome: The system maintains proper dynamic response and waveform quality.

Check Sensor Leveling

Incorrect leveling can cause inaccurate pressure values even when the waveform shape appears normal.

Action:

Expected outcome: The numeric pressure value becomes more accurate after proper leveling.

Why it matters: A waveform can look present but still display the wrong pressure if the transducer is not level.

Check Zeroing

An unzeroed or incorrectly zeroed pressure channel can create inaccurate values or misleading baseline pressure.

Action:

Expected outcome: The waveform baseline and numeric pressure are corrected.

If the issue resolves after proper zeroing, document the zeroing issue and stop.

Check Pressure Type and Channel Assignment

Wrong pressure type selection or channel assignment can cause confusion, incorrect labeling, or incorrect interpretation.

Action:

Expected outcome: The displayed waveform and numeric value match the connected invasive pressure source.

Compare Against Another Monitor or Known-Good Setup

Use substitution to separate monitor/cable failure from patient line or disposable setup issues.

Action:

Expected outcome: The likely failed component is identified without internal disassembly.

Check for Display, Freeze, or Software Behavior

If the waveform appears frozen, delayed, or intermittently missing, the issue may not be the pressure line itself.

Action:

Expected outcome: Temporary display or software-related behavior is ruled out.

Inspect for Physical Damage or Fluid Intrusion

Hemodynamic monitoring equipment is used around fluids, pressure bags, and invasive line setups.

Action:

Expected outcome: Unsafe or unreliable accessories are removed before further patient use.

If the Problem Persists

If the invasive pressure waveform remains absent, dampened, inaccurate, intermittent, or unreliable after external setup, cable, sensor, leveling, zeroing, and configuration checks have been completed, the issue may involve an internal monitor fault, pressure cable electronics issue, input port failure, or software-related problem.

The device should be removed from service, labeled Out of Service, sent for repair or bench evaluation, and tested with known-good pressure cables and compatible pressure simulation equipment if available.

Knowing when to stop is proper troubleshooting. Do not keep troubleshooting invasive pressure monitoring on an active patient when the reading cannot be trusted.

Clinical Use Tip

Do not troubleshoot an invasive pressure waveform on an active patient unless clinical staff confirms it is safe. If the pressure value is clinically important, move the patient to a reliable monitoring setup first, then evaluate the HemoSphere system separately. This helps protect therapy continuity and avoids clinical decisions based on an untrusted pressure value.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported the HemoSphere monitor had no arterial pressure waveform and displayed unreliable invasive pressure values during setup."

Cause

What was observed during troubleshooting.

Example:
"Found the pressure transducer cable connection loose at the HemoSphere pressure cable, causing intermittent waveform detection."

Resolution

What action was taken.

Example:
"Reseated the pressure cable connection, verified stable invasive pressure waveform with clinical staff, confirmed proper leveling and zeroing, and returned the monitor to service."

Helpful Details to Include (If Known)

Final Thought

Invasive pressure troubleshooting should stay patient-safety focused and logical. Start with the line, sensor, cable, leveling, zeroing, and setup before suspecting the monitor. If the pressure signal cannot be trusted after basic checks, remove the device from service and document the complaint, cause, and resolution clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide