Network / Export / EMR Connectivity Failure

On this page

Asset Type

Hemodynamic Monitor

Manufacturer

Edwards Lifesciences

Model

HemoSphere Advanced Monitoring System

What This Guide Helps With

Troubleshooting HemoSphere network, data export, EMR, interface, or connectivity failures caused by setup, cabling, network, or configuration issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the HemoSphere monitor is not the only source being used for active patient decision-making before troubleshooting connectivity.

Action:

Expected outcome: Patient monitoring continues safely even if EMR export or network communication is unavailable.

Why it matters: A connectivity issue is usually secondary to the primary monitoring function. Patient safety and continued visibility of hemodynamic values come first.

Verify the Reported Connectivity Problem

Confirm exactly what is failing before changing settings.

Action:

Expected outcome: The failure is narrowed to a single monitor, patient association issue, network issue, or larger interface/EMR issue.

Why it matters: One device failing points toward local cabling/configuration. Multiple devices failing points toward network, interface engine, EMR, or server-side problems.

Check the Network Cable and Physical Connection

Start with the simplest external connection.

Action:

Expected outcome: The monitor has a secure physical network connection.

Why it matters: A loose or damaged Ethernet cable can cause complete loss of export, intermittent communication, or delayed data transmission.

If this resolves the issue, return the device to service and document the cable or connection problem.

Verify the Network Jack or Drop

Confirm the wall jack or network drop is active.

Action:

Expected outcome: The monitor is connected to an active, correct network port.

Why it matters: Network/export failures are often caused by the device being connected to the wrong jack, an inactive port, or a port assigned to the wrong network/VLAN.

If the jack is inactive or incorrectly configured, escalate to IT/network support.

Confirm the Correct Patient and Bed Association

Verify the monitor is associated with the correct patient/location workflow.

Action:

Expected outcome: The device, patient, and bed/location match the expected EMR destination.

Why it matters: The monitor may be working correctly but exporting data to the wrong patient context, wrong location, or not exporting because patient association is incomplete.

If this resolves the issue, confirm values are crossing correctly before closing the work order.

Check Network Status or Connectivity Indicators

Review available system status information without making deep configuration changes.

Action:

Expected outcome: The monitor provides a clear indication of whether it is connected, disconnected, or unable to communicate with a destination system.

Why it matters: Exact messages help separate physical network failure from interface, server, configuration, or EMR-side issues.

Power Cycle Only if Clinically Safe

Restart the monitor only when it will not interrupt active patient monitoring.

Action:

Expected outcome: Temporary software or communication lockups clear after restart.

Why it matters: Some communication failures are caused by temporary software, network handshake, or session problems. A safe restart can restore function without repair.

If this resolves the issue, monitor briefly to confirm values continue exporting.

Compare With Another Known-Good Monitor

Determine whether the issue follows the device or the network location.

Action:

Expected outcome:

Why it matters: This is one of the best ways to separate a device problem from a network infrastructure problem.

Check for Recent Changes

Look for timing clues.

Action:

Expected outcome: A recent change may explain why communication stopped.

Why it matters: Connectivity failures often happen after network changes, interface engine work, room moves, or patient association changes rather than hardware failure.

Escalate to IT / Interface Support When Appropriate

Involve the correct support team once local device checks are complete.

Action:

Expected outcome: Network, interface, or EMR-side causes are investigated by the responsible team.

Why it matters: Clinical Engineering can verify the device and external setup, but EMR routing, network switch configuration, firewall rules, and interface engine problems usually require IT or informatics support.

Verify Restoration Before Closing

Confirm that the repair or correction actually restored data flow.

Action:

Expected outcome: Data is flowing to the correct destination and clinical staff confirm normal operation.

Why it matters: Network communication may appear restored at the device but still fail at the EMR or patient association level.

If the Problem Persists

If the HemoSphere monitor continues to fail network/export/EMR communication after cable, jack, patient association, restart, and comparison testing have been completed, common external causes have been ruled out.

The device should be:

Knowing when to stop is proper troubleshooting. Clinical Engineering should not continue guessing at network or EMR-side settings without the correct access, documentation, and support path.

Clinical Use Tip

Do not troubleshoot network/export issues in a way that interrupts active patient monitoring. If the monitor is being used for live hemodynamic management, have clinical staff continue monitoring locally or move the patient to another safe monitoring setup before restarting, moving, or disconnecting the device.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Edwards HemoSphere monitor in OR 4 was displaying values locally but not exporting hemodynamic data to the EMR."

Cause

What was observed during troubleshooting.

Example:
"Found Ethernet connection secure and monitor operating normally, but network drop did not provide active clinical network connectivity."

Resolution

What action was taken.

Example:
"Verified monitor operation on a known-good network jack, escalated inactive wall port to IT, and returned the monitor to service once EMR data export was confirmed."

Helpful Details to Include (If Known)

Final Thought

Network and EMR issues should be approached logically. Confirm patient safety first, then check the physical connection, network jack, patient association, device status, and scope of the failure. Once the local device and external setup are ruled out, escalation to IT, informatics, interface support, or repair is the correct next step. Good documentation protects the patient, the department, and the troubleshooting process.

That is successful troubleshooting.