Eitan Medical Sapphire

Administration Set, Door, or Cassette Loading Failure

On this page

Asset Type

Infusion Pump

Manufacturer

Eitan Medical

Model

Sapphire

What This Guide Helps With

Troubleshoots loading failures caused by incorrect set placement, damaged disposables, obstructed door movement, latch problems, or contaminated loading surfaces.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Therapy

Do not continue patient therapy on a pump that cannot reliably secure or recognize the administration set.

Transfer the infusion to another verified device if therapy is active.

Expected outcome: The patient remains on reliable therapy and the affected pump can be inspected safely.

2. Confirm the Reported Failure

Determine whether the issue is:

Expected outcome: The exact mechanical or recognition problem is identified. If correct loading resolves the issue immediately and passes verification, troubleshooting can stop.

3. Inspect the Administration Set

Verify that the set being used is compatible with the pump and that its loading components are not bent, cracked, distorted, contaminated, or incorrectly assembled.

Do not force an incompatible disposable into the pump.

Expected outcome: A compatible, undamaged set is available. If replacing a damaged or incorrect set resolves the problem, troubleshooting can stop.

4. Inspect the Loading Path

With the pump removed from patient use, inspect accessible loading surfaces for:

Clean only according to approved procedures.

Expected outcome: The loading path is clean, unobstructed, and dry. If the set loads normally afterward, troubleshooting can stop.

5. Reload the Set Carefully

Unload and reinstall the set according to approved procedures.

Ensure all designated tubing sections are seated fully and are not twisted or protruding into the door path.

Expected outcome: The administration set lies correctly in the intended path and the door closes normally. If the pump recognizes the set and functions normally, troubleshooting can stop.

6. Inspect Door and Latch Movement

Open and close the door while the pump is off the patient.

Look for:

Do not bend or reshape latch components.

Expected outcome: The door moves smoothly and latches securely without excessive force. If visible structural damage is present, remove the pump from service.

7. Compare With a Known-Good Set

Use a known-good compatible administration set to determine whether the problem follows the disposable or remains with the pump.

Expected outcome: The known-good set loads and is recognized normally. If so, the original set is the likely cause and troubleshooting can stop.

8. Verify Pump Recognition

Power the pump and verify that the correctly installed set is accepted and that no loading-related warning remains.

Do not bypass set-recognition or door-detection features.

Expected outcome: The pump transitions to normal ready status with the set properly installed. If so, proceed to final verification.

9. Perform Final Functional Verification

Using an approved bench setup, verify startup, door operation, administration-set loading, infusion initiation, alarms, and general pump operation.

Cycle the door and set installation more than once if the original issue was intermittent.

Expected outcome: Loading and recognition are repeatable and reliable. If all required testing passes, troubleshooting can stop and the pump may be returned to service.

10. Escalate Mechanical or Recognition Failure

If a known-good compatible set cannot be loaded, recognized, or securely retained, stop external troubleshooting.

Do not disassemble the door mechanism, internal sensors, pumping assembly, or latch system unless authorized and qualified to do so.

Expected outcome: The pump is removed from service for proper bench evaluation.

If the Problem Persists

External causes such as incorrect disposable selection, misloading, contamination, tubing obstruction, and obvious accessory damage have been ruled out. Remaining causes may involve the door mechanism, set-detection system, pumping interface, alignment, or internal service-level components.

The device should be:

After repair, verify secure loading, recognition, alarm function, and infusion performance before return to clinical use.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Never force the door closed around a poorly seated administration set; doing so can damage the set, pump, or fluid pathway.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported the Sapphire door would not close after installing a new administration set."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the tubing routed outside the designated loading path and caught beneath the door."

Resolution

What action was taken.

Example:
"Reloaded the administration set correctly and verified repeated door closure, set recognition, and normal pump operation."

Helpful Details to Include (If Known)

Final Thought

Loading problems should be approached as external mechanical and disposable issues first. Verify the set, routing, cleanliness, and door operation before suspecting an internal recognition or latch failure.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide