Eitan Medical Sapphire

Air-in-Line Alarm or Repeated False Air Detection

On this page

Asset Type

Infusion Pump

Manufacturer

Eitan Medical

Model

Sapphire

What This Guide Helps With

Troubleshoots air-in-line alarms caused by actual air, poor priming, tubing placement, damaged sets, contamination, or suspected false air detection.

Step-by-Step Troubleshooting

1. Protect the Patient and Stop Unsafe Delivery

Do not continue an infusion through tubing containing visible air when doing so may pose a patient risk.

Move the patient to another verified infusion device or otherwise maintain prescribed therapy according to clinical policy before testing the affected pump.

Expected outcome: The patient is protected from unintended air delivery and therapy continuity is maintained.

2. Confirm the Exact Alarm Condition

Verify the pump's displayed alarm and determine whether visible air is actually present in the tubing.

Ask whether the alarm occurs at startup, after a container change, during transport, or repeatedly with no visible bubbles.

Expected outcome: The problem is categorized as actual air introduction or suspected false detection. If the pump operates normally after correcting a clearly visible air condition, troubleshooting can stop after verification.

3. Inspect the Entire Fluid Path for Air

Trace the administration set from the source container through the pump and downstream tubing.

Look for:

Expected outcome: No clinically significant visible air remains in the test fluid path. If removing the actual air resolves the alarm, troubleshooting can stop.

4. Verify Proper Priming

Replace or reprime the administration set according to approved clinical procedures if air is present.

Clinical Engineering should not improvise priming methods or bypass air-detection functions.

Expected outcome: The administration set is completely and properly primed. If the pump operates normally after correct priming, troubleshooting can stop.

5. Inspect Tubing at the Air-Detection Area

Check the tubing where it passes through the pump's air-detection region.

Verify that it is:

Expected outcome: The tubing is seated consistently in the sensing area. If reseating the set eliminates false alarms, troubleshooting can stop.

6. Inspect the Administration Set

Examine the disposable for clouding, damage, unusual wall thickness, deformation, contamination, or other visible abnormalities.

If there is any doubt, replace it with a known-good compatible administration set.

Expected outcome: A known-good compatible set is installed. If the problem does not recur, the original set was the likely cause and troubleshooting can stop.

7. Clean Accessible External Sensing Surfaces

If permitted by approved cleaning instructions, inspect and clean externally accessible sensor areas for residue, dried fluid, dust, or contamination.

Do not introduce fluid into the pump or disassemble the sensor assembly.

Expected outcome: External sensing surfaces are clean and dry. If false alarms stop after approved cleaning, troubleshooting can stop after functional verification.

8. Check Environmental Factors

Verify that condensation, extreme lighting, contamination, or unusual tubing positioning is not associated with the alarm.

Test in a normal controlled environment with a known-good set.

Expected outcome: Air detection is stable under normal operating conditions. If environmental correction resolves the issue, troubleshooting can stop.

9. Perform Controlled Functional Verification

Using an approved test setup and known-good compatible set, verify normal infusion operation and appropriate alarm behavior.

Do not deliberately introduce unsafe amounts of air into a patient-connected line.

Expected outcome: The pump operates without false air alarms and responds appropriately during approved testing. If all checks pass, troubleshooting can stop and the pump may be returned to service according to policy.

10. Escalate Persistent False Detection

If repeated air-in-line alarms continue with fully primed tubing, a known-good set, and clean external sensing surfaces, stop troubleshooting.

Do not disable, bypass, adjust, or calibrate the air detector without authorized manufacturer procedures.

Expected outcome: The pump is removed from clinical use for service-level evaluation.

If the Problem Persists

Common external causes such as actual air, inadequate priming, tubing misplacement, set defects, and contamination have been ruled out. Remaining causes may involve the air-detection system, configuration, calibration, or internal electronics.

The device should be:

Complete applicable alarm and infusion-performance testing before return to service.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Never dismiss repeated air alarms as nuisance alarms until visible air, priming quality, tubing placement, and disposable condition have all been verified.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported repeated air-in-line alarms even though no obvious air was visible during use."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the administration tubing twisted and incompletely seated in the pump's sensing path."

Resolution

What action was taken.

Example:
"Reloaded the tubing correctly, verified the set was fully primed, and completed functional testing with no further false air alarms."

Helpful Details to Include (If Known)

Final Thought

Air alarms require both patient-safety awareness and careful external verification. Confirm real air, proper priming, tubing placement, and disposable condition before suspecting the sensing system, and escalate persistent false detection appropriately.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide