Eitan Medical Sapphire

Battery Will Not Charge or Runtime Is Short

On this page

Asset Type

Infusion Pump

Manufacturer

Eitan Medical

Model

Sapphire

What This Guide Helps With

Troubleshoots charging failure or short runtime caused by power supply issues, poor connections, charging conditions, battery age, or suspected battery-system failure.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Therapy

If battery performance is unreliable during active therapy or transport, move the patient to another verified pump or ensure dependable AC operation according to clinical requirements.

Do not rely on a pump with uncertain battery runtime for transport or power-interruption protection.

Expected outcome: Patient therapy no longer depends on the questionable battery system.

2. Confirm the Battery Complaint

Determine whether the problem is:

Expected outcome: The battery failure mode is clearly defined. If the issue is only an incorrect external connection and charging resumes normally, troubleshooting can stop after verification.

3. Verify AC Power Source

Confirm that the receptacle is energized and appropriate for use.

If needed, test the outlet or use a known-good verified outlet.

Expected outcome: Reliable AC power is available. If changing to a functional outlet restores charging, troubleshooting can stop.

4. Inspect the Approved Power Supply

Inspect the adapter, cord, strain relief, plug, and pump-side connector for:

Do not use damaged power accessories.

Expected outcome: The power supply and connector are physically intact. If replacing a damaged external power accessory resolves charging, troubleshooting can stop.

5. Confirm Power Connection

Connect the approved power supply securely and observe the pump's charging or external-power indication.

Gently verify that ordinary connector movement does not interrupt charging.

Expected outcome: External power is recognized consistently. If reseating the connection resolves the issue and charging remains stable, troubleshooting can stop.

6. Compare With a Known-Good Power Supply

Where compatible and permitted, use a known-good approved power supply to determine whether the fault follows the adapter or remains with the pump.

Expected outcome: The pump charges normally with the known-good supply. If so, the original external power supply is the likely cause and troubleshooting can stop.

7. Allow Normal Charging Opportunity

If the battery was deeply discharged, allow the pump to remain connected to verified AC power under normal approved charging conditions.

Do not assume immediate full charge.

Expected outcome: Battery status increases normally over time. If charging proceeds normally and subsequent runtime is acceptable under approved testing, troubleshooting can stop.

8. Assess Battery Runtime

After appropriate charging, operate the pump using an approved bench setup while monitoring battery behavior.

Use facility or manufacturer-approved battery evaluation criteria rather than invented runtime thresholds.

Expected outcome: Battery operation remains stable for the approved evaluation. If runtime is inadequate, continue to escalation or qualified battery replacement procedures.

9. Perform Final Functional Verification

After correction or qualified battery service, verify:

Expected outcome: The pump operates reliably on both approved external power and battery. If all required checks pass, troubleshooting can stop and the device may be returned to service.

10. Escalate Persistent Battery Problems

If charging does not occur with verified AC power and a known-good approved adapter, or if runtime remains inadequate after proper charging, remove the device from service.

Do not open the pump or replace internal battery components unless trained and authorized under approved service procedures.

Expected outcome: A likely battery, charging-circuit, connector, or power-management problem is referred for service.

If the Problem Persists

External power, receptacle, adapter, and connector problems have been ruled out. Remaining causes may involve battery degradation, internal charging circuitry, battery communication, power-management components, or configuration.

The device should be:

Complete charging, battery-operation, and full functional verification before return to clinical service.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Do not use a pump with uncertain battery capacity for patient transport, even if it appears normal while connected to AC power.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported the Sapphire would not charge and shut down shortly after being disconnected from AC power."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external power adapter connector was damaged and intermittently losing contact."

Resolution

What action was taken.

Example:
"Replaced the damaged approved power adapter and verified stable charging, battery operation, and normal pump function."

Helpful Details to Include (If Known)

Final Thought

Battery complaints should be separated into external-power, charging, and capacity problems. Verify the outlet and adapter first, then evaluate battery performance using approved methods before escalating internal power concerns.

That is successful troubleshooting.