Eitan Medical Sapphire

Downstream Occlusion or High-Pressure Alarm

On this page

Asset Type

Infusion Pump

Manufacturer

Eitan Medical

Model

Sapphire

What This Guide Helps With

Troubleshoots repeated downstream occlusion or high-pressure alarms caused by tubing restrictions, clamps, access devices, positioning, or external infusion-path problems.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Therapy

Do not troubleshoot an unreliable infusion pump while a patient depends on it. If therapy is active or clinically time-sensitive, transfer the infusion to another verified device or alternate delivery method according to clinical policy.

Inspect the pump and administration set for leakage, cracking, damage, contamination, or evidence of excessive pressure.

Expected outcome: The patient remains supported by reliable therapy, and the affected pump can be evaluated safely. If unsafe conditions are present, remove the pump from service and stop troubleshooting.

2. Confirm the Reported Alarm

Review the displayed alarm or staff report and confirm whether the condition is consistently identified as downstream occlusion or high pressure.

Ask whether the alarm occurred immediately after setup, after patient movement, during medication administration, or only with a specific access site or administration set.

Expected outcome: The reported condition is reproduced or clearly characterized. If the alarm cannot be reproduced and the pump passes functional verification, troubleshooting can stop after documentation.

3. Inspect the Downstream Tubing Path

Trace the administration set from the pump toward the patient or test load. Look for:

Correct any obvious restriction without changing therapy settings.

Expected outcome: The downstream pathway is open and unrestricted. If correcting the restriction clears the alarm and normal operation is verified, troubleshooting can stop.

4. Check Patient-Side or Test-Side Resistance

If clinically appropriate and after the patient has been moved to alternate therapy, determine whether the problem follows the pump or the downstream access path.

Use an approved test setup or known-good external fluid path rather than attempting to force fluid through a questionable patient access device.

Expected outcome: The pump operates normally through a verified unrestricted pathway. If so, the original downstream access path or accessory is the likely cause and troubleshooting can stop after documentation.

5. Inspect the Administration Set

Verify that the installed administration set is compatible with the pump and is not damaged, collapsed, twisted, stretched, contaminated, or improperly seated.

Pay particular attention to tubing immediately where it exits the pump, since pinching or improper routing can create resistance.

Expected outcome: The administration set is correctly installed and free of visible restriction. If replacing a questionable set with a known-good compatible set resolves the alarm, troubleshooting can stop.

6. Verify Door and Set Loading

Open and reload the administration set according to approved operating procedures. Confirm the tubing is fully seated in the intended path and that the door closes and latches without excessive force.

Do not defeat door or set-detection mechanisms.

Expected outcome: The set is correctly loaded, the door closes normally, and the pump recognizes the setup. If the alarm no longer occurs during testing, troubleshooting can stop.

7. Check Positioning and External Pressure

Verify that the tubing is not being compressed by the pump mount, pole clamp, bed movement, transport equipment, or nearby accessories.

If the reported alarm is position-dependent, gently reposition external tubing and mounting components while observing operation.

Expected outcome: Normal operation is maintained regardless of ordinary equipment positioning. If eliminating an external pinch point resolves the issue, troubleshooting can stop.

8. Test With a Known-Good Setup

Using a known-good compatible administration set and appropriate test fluid path, operate the pump under controlled conditions.

Do not use a patient as the test load.

Expected outcome: The pump infuses without generating an inappropriate downstream occlusion or high-pressure alarm. If successful, the pump may proceed to final verification.

9. Perform Final Functional Verification

Verify normal startup, programming, infusion initiation, alarm annunciation, display function, and expected flow behavior using approved Clinical Engineering test methods and equipment.

Inspect for abnormal noise, heat, odor, leakage, or intermittent alarms.

Expected outcome: The pump operates reliably and no false high-pressure condition occurs. If all required checks pass, troubleshooting can stop and the device may be returned to service according to facility policy.

10. Escalate if the Alarm Persists

If the pump continues to report downstream occlusion or high pressure with a verified administration set and unrestricted test pathway, stop external troubleshooting.

Do not perform unauthorized internal pressure-sensor adjustment, calibration, board-level repair, or disassembly.

Expected outcome: A persistent device-related fault is identified as requiring qualified service evaluation.

If the Problem Persists

Common external restrictions, tubing problems, loading errors, and patient-side causes have been ruled out. Remaining possibilities may involve the pump's pressure-sensing system, pumping mechanism, configuration, calibration status, or another internal service-level condition.

The device should be:

After corrective service, complete applicable operational, alarm, and infusion-performance verification before returning the pump to clinical use.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Move the patient to another verified infusion device before investigating repeated pressure alarms so therapy is not interrupted during testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported repeated downstream occlusion alarms during infusion despite restarting the Sapphire pump."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the downstream tubing pinched beneath the pump mounting hardware."

Resolution

What action was taken.

Example:
"Repositioned the tubing, verified an unrestricted fluid path, and completed functional testing with no further occlusion alarms."

Helpful Details to Include (If Known)

Final Thought

Protect the patient first, then work systematically from the external infusion pathway toward the pump. Verify tubing, accessories, loading, and setup before assuming an internal pressure-sensing problem, and document both the cause and final verification clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide