Eitan Medical Sapphire

Upstream Occlusion or Empty-Container Alarm

On this page

Asset Type

Infusion Pump

Manufacturer

Eitan Medical

Model

Sapphire

What This Guide Helps With

Troubleshoots upstream occlusion or empty-container alarms caused by depleted fluid, closed clamps, tubing restrictions, container position, or administration-set problems.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Therapy

Do not troubleshoot an unreliable infusion pump while a patient depends on it. Transfer clinically necessary therapy to another verified pump or approved delivery method before extended testing.

Inspect for an actually empty medication container, interruption of critical therapy, leakage, or damaged tubing.

Expected outcome: Therapy continuity is maintained and the pump can be tested safely. If the pump or set is damaged or leaking, remove it from service immediately.

2. Confirm the Exact Report

Verify whether the pump displays an upstream occlusion, empty-container, or similar no-fluid condition.

Determine whether the alarm occurs immediately after starting, near the end of the container, after repositioning, or only with a particular administration set.

Expected outcome: The failure pattern is clearly defined. If the issue cannot be reproduced and the pump completes functional testing normally, troubleshooting can stop.

3. Verify Fluid Is Available

Confirm that the source container contains sufficient fluid and that the tubing inlet remains submerged or properly connected as applicable.

Inspect for a collapsed flexible bag, improperly vented rigid container when applicable, or a source container positioned in a way that prevents normal fluid delivery.

Expected outcome: An adequate fluid source is present and able to supply the pump. If correcting the container condition resolves the alarm, troubleshooting can stop.

4. Check Upstream Clamps and Connections

Trace the administration set from the source container to the pump.

Verify:

Expected outcome: The entire upstream path is open. If correcting a clamp or connection clears the alarm, troubleshooting can stop.

5. Inspect the Upstream Tubing

Look for kinks, collapsed tubing, twisting, stretching, tubing trapped beneath equipment, or tight bends near the pump.

Pay special attention to tubing passing through pole-mounted accessories.

Expected outcome: Tubing remains open and unrestricted throughout normal positioning. If eliminating a kink resolves the issue, troubleshooting can stop.

6. Verify Administration Set Installation

Unload and reinstall the compatible administration set using approved procedures.

Confirm that the set is routed correctly and not caught, pinched, partially seated, or distorted where it enters the pumping mechanism.

Expected outcome: The set is correctly installed and recognized. If the pump operates normally after proper reloading, troubleshooting can stop.

7. Substitute a Known-Good Set and Fluid Source

When the cause is not obvious, install a known-good compatible administration set and suitable test fluid container.

Do not continue using a suspect disposable during verification.

Expected outcome: The pump runs without an inappropriate upstream or empty-container alarm. If so, the removed administration set or fluid source was the likely cause and troubleshooting can stop.

8. Evaluate Environmental and Positioning Factors

Confirm that ordinary movement of the IV pole, pump, container, or tubing does not create an upstream restriction.

Check whether the fluid container is hanging freely and whether accessories are compressing the tubing.

Expected outcome: Stable operation continues through normal equipment positioning. If correcting positioning resolves the issue, troubleshooting can stop.

9. Perform Final Functional Verification

Using an approved bench setup, verify startup, set recognition, infusion initiation, alarm operation, display function, and stable delivery through a known-good fluid path.

Expected outcome: The pump operates normally without false upstream or empty-container alarms. If all required tests pass, troubleshooting can stop and the pump may be returned to service according to policy.

10. Escalate if the Alarm Persists

If the alarm remains with a full test container, unrestricted tubing, and a known-good compatible set, discontinue external troubleshooting.

Do not adjust internal sensors or enter unauthorized service or calibration functions.

Expected outcome: The pump is identified as requiring service-level evaluation.

If the Problem Persists

External causes such as empty containers, clamps, tubing restrictions, improper loading, and disposable-set problems have been ruled out. Remaining causes may involve sensing, pumping performance, configuration, calibration, or another internal service-level condition.

The device should be:

Complete appropriate return-to-service testing before clinical use.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Do not assume an empty-container alarm is false until the source container and entire upstream fluid path have been physically verified.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported the Sapphire repeatedly alarmed for upstream occlusion shortly after starting an infusion."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the upstream roller clamp partially closed on the administration set."

Resolution

What action was taken.

Example:
"Opened the clamp, verified unrestricted flow, and completed pump functional testing without recurrence."

Helpful Details to Include (If Known)

Final Thought

Verify the simple fluid-source and tubing conditions first. An upstream alarm often reflects a real restriction outside the pump, and careful verification should precede any assumption of internal failure or escalation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide